background image

CE WT-BA-e-2010.docx 

 

Brief description 

 

Analogue weighing transmitter 

 

Power supply 12V or 24V 

 

Output signal voltage or current 

 

E.g. for transmission to SPS, analog measuring card, etc. 

 

Integrated overvoltage protection 

 

Reverse polarity protection at the input and protection of the output 

Available models 

Type/part number 

Type 

Series 

Sensor 

connections 

Supply 

voltage 

Output signal 

Housing 

Protection 

class 

CE WT1-Y1 

CE 
WTY1 

1  12V 

Analog (4 - 20 
mA) 

Sheet Steel    

CE WT2-Y1 

CE 
WTY1 

1  24V  

Analog (4 - 20 
mA) 

Sheet Steel    

CE WT3-Y1 

CE 
WTY1 

1  12V 

Analog 0 +/-5V  Sheet Steel    

CE WT4-Y1 

CE 
WTY1 

1  24V  

Analog 0 +/-5V  Sheet Steel    

CE WT1-Y2 

CE 
WTY2 

1  12V 

Analog (4 - 20 
mA) 

Aluminium  IP65 

CE WT2-Y2 

CE 
WTY2 

1  24V  

Analog (4 - 20 
mA) 

Aluminium  IP65 

CE WT1-Y4 

CE 
WTY4 

4  12V 

Analog (4 - 20 
mA) 

Aluminium  IP65 

CE WT2-Y4 

CE 
WTY4 

4  24V  

Analog (4 - 20 
mA) 

Aluminium  IP65 

 

Dimensions 

Type Series 

In mm 

CE WTY1 

112x45x33 

CE WTY2 

110x64x37 

CE WTY4 

158x100x65 

 

Check before use  

After receipt of the device, it should be checked in advance that no transport damage 
has occurred,  that  the  outer  packaging,  the plastic  housing,  other parts  or  even  the 
device itself have not been damaged. If any damage is evident, please notify SAUTER 
GmbH immediately. 

Summary of Contents for CE WT1-Y1

Page 1: ...Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu User Manual Data Sheet SAUTER CE WTY1 CE CT Y2 CE WT Y4 V 1 0 11 2020 GB CE WT BA e 2010 docx ...

Page 2: ... Summarize 1 Brief description 3 2 Available models 3 3 Dimensions 3 4 Check before use 3 5 Connection 4 5 1 Sensor 4 5 1 1 CE WTY1 and CE WTY2 4 5 1 2 CE WTY4 4 5 2 Power supply and signal 5 5 3 Adjust zero point 5 5 3 1 CE WTY1 and CE WTY2 5 5 4 Settings CE WTY4 6 6 Declaration of conformity 6 ...

Page 3: ...mA Sheet Steel CE WT3 Y1 CE WTY1 1 12V Analog 0 5V Sheet Steel CE WT4 Y1 CE WTY1 1 24V Analog 0 5V Sheet Steel CE WT1 Y2 CE WTY2 1 12V Analog 4 20 mA Aluminium IP65 CE WT2 Y2 CE WTY2 1 24V Analog 4 20 mA Aluminium IP65 CE WT1 Y4 CE WTY4 4 12V Analog 4 20 mA Aluminium IP65 CE WT2 Y4 CE WTY4 4 24V Analog 4 20 mA Aluminium IP65 3 Dimensions Type Series In mm CE WTY1 112x45x33 CE WTY2 110x64x37 CE WTY...

Page 4: ...WTY1 and CE WTY2 Connection to connector with 5 pins Pin 1 IN Input 2 IN Input 3 EX Excitation 4 EX Excitation 5 Shield 5 1 2 CE WTY4 Sx Input Sx Input Ex Excitation Ex Excitation PB Shield The pins are numbered on the front Connect the 5 wires of the sensor ...

Page 5: ...mitter if it displays incorrectly Open the transmitter and find the only potentiometer without silica gel The zero adjustment of the transmitter could be completed by adjusting the 5K potentiometer Please note that all 100Ohm potentiometers are sealed with silica gel which means that the user was NOT allowed to make any adjustments to them Please DO NOT adjust any components of the product as Conn...

Page 6: ... can set the current zero point of the analog output W6 can set the gain of the circuit in the voltage output 0 5V 0 10V W8 can set the amplification of the circuit in the current output 4 20mA Note Before delivery all transmitters are calibrated without special need do not adjust the potentiometer 6 Declaration of conformity To view the CE declaration please click on the following link https www ...