Set the parameters of the analysis

Set the parameters of the analysis 19

7. Set the parameters of the analysis

This menu allows to set the following itemes regarding the analysis:

• Fuel

• Pump after autozero

• CO pump cut-off level

• Zero draft sensor

• Reference O

2

• Atmospheric pressure

• NOx factor

• Air temperature

• Smoke/soot

• Display configuration

• Alarms

• Stack cross-sectional area

7.1 Set the fuel

This part allows to select the required fuels for the gas analysis but it also allows to add customized fuels.

7.1.1 Select the fuel

The default list of fuels is dependent on the country selected. Possible fuels listed include the following:

• Natural Gas (H)

• Biogas

• Natural Gas (L) Groningue

• Firewood 20% (Soft)

• Natural Gas North America

• Firewood 50% (Soft)

• #2 Oil / Domestic

• Firewood 20% (Hard)

• #3 Oil / Heavy

• Firewood 50% (Hard)

• #6 Oil / RFO

• Wood Chips / Pellets 8%

• Propane

• Cannel Coal

• Butane

• Low Volatile Coal

• LPG

• Hog Fuel

• Biofuel 5%

• Peat

• Diesel

• Biomass

• Coke Gas

“Settings > Analysis”

 screen is displayed.

• Tap 

“Fuel”

.

• Select the required fuel.

To see features of the fuel, tap on it. The following information about the fuel is displayed: A1, B, K, CO

2

t(%), V 

dry gas (m

3

/ucb), M air (kg/ucb), MH

2

O (kg/ucb) and PCI (MJ/ucb)

7.1.2 Create a fuel

It is possible to create a customized fuel.

“Settings > Analysis”

 screen is displayed.

• Tap 

“Fuel”

.

• Tap 

"Add fuel"

 on the bottom of the screen.

• Name the fuel and validate it tapping 
• Enter the following features of the fuel: A1, B, K, CO

2

t(%), V dry gas (m

3

/ucb), M air (kg/ucb), MH

2

O (kg/ucb) and 

PCI (MJ/ucb).

• Tap 

"Save"

.

The customized fuel is then integrated at the end of the fuel list.

If a very specific fuel must be created please contact the Sauermann Service Center for further information.

A customized fuel can be deleted: 

• Tap the name of the customized fuel on the fuel list.

Features of this fuel are displayed.

• Tap 

 on the top of the screen.

A message is displayed asking to confirm the deleting.

• Tap Yes to confirm.

7.2 Set the CO pump cut-off level

This part allows to define the CO pump cut-off level.

“Settings > Analysis”

 screen is displayed.

• Tap 

“CO pump cut-off level”

.

• Enter the required level between 1000 and 8000 ppm.

To enable the CO pump cut-off level, activate the protection at the top of the screen.

In normal operation it is recommended to set the CO pump cut-off level threshold higher than 
the CO dilution threshold.

• 

Summary of Contents for SI-CA 130

Page 1: ...www sauermanngroup com USER MANUAL Si CA 130 GAS ANALYZER...

Page 2: ...s results 14 5 8 Use the Zoom function 15 6 Set the general parameters of the analyzer 16 6 1 Set the time 16 6 2 Set the time zone 16 6 3 Set the date 16 6 4 Set the language 16 6 5 Set the country 1...

Page 3: ...fingers in movable zones of the device articulations The device must not be used in ATEX zones according to applicable standards Do not store the device with solvents Do not use desiccants Do not use...

Page 4: ...to maintain the contact tip inside the duct The probe is connected to the instrument by 3 m 10 of dual hose Other probe lengths 75 cm 1 m and 1 5 m are avalaible as options Interchangeable measuremen...

Page 5: ...asurement also for Heat Exchanger Test NO Direct NO sensor measurement NO L Direct Low NO sensor measurement NOx Calculated NOx value NOx L Calculated Low NOx value with NOx factor CO2 Calculated CO2...

Page 6: ...lor Touch Screen with Graphing Size 480 x 272 pixels Keypad 1 On Off key Material ABS PC Protection IP42 Connection Wireless class 2 range range frequency from 2402 MHz to 2480 MHz with a transmit pow...

Page 7: ...Gas Analysis screen The screen displays values of the different parameters of the analysis Perform measurement for at least 3 minutes until measurements stabilize 5 3 Perform a draft measurement Duri...

Page 8: ...arameters to display Select the time period to display Tap Start The analyzer displays the measurements as graph 5 6 Save the analysis During the combustion gas analysis when measurements are constant...

Page 9: ...e Settings General screen is displayed Tap Language Select the language among English French Spanish Italian German Portuguese and Chinese 6 5 Set the country Settings General screen is displayed Tap...

Page 10: ...following information about the fuel is displayed A1 B K CO2 t V dry gas m3 ucb M air kg ucb MH2 O kg ucb and PCI MJ ucb 7 1 2 Create a fuel It is possible to create a customized fuel Settings Analys...

Page 11: ...or falling the alarm of the instrument is activated when the measured value is below the defined threshold A maximum of 5 different alarms can be set Settings Analysis screen is displayed Tap Alarms...

Page 12: ...ernal NTC sensor of the analyzer if no probe is connected and if no value entered To enter an air temperature Settings Analysis screen is displayed Tap Air temperature Activate it at the top of the sc...

Page 13: ...e the data saving mode manual or data logger Settings screen is displayed Tap Data saving Select Manual data will be manualy saved during measurement or Select Data logger data will be saved according...

Page 14: ...rinted default number is 1 Settings Printer screen is displayed Tap Number of copies Use up and down arrows to define the number of copies Maximum number of copies is five 11 Analyzer information This...

Page 15: ...12 Perform other measurements The analyzer can perform other measurements than a gas analysis CO monitor Gas tightness test Gas pump flow rate Heat exchanger integrity test Stack gas velocity 12 1 Pe...

Page 16: ...or combustion chamber The pump of the analyzer must be on to perform this test Other measurements screen is displayed Tap Heat Exchanger Integrity Test Insert the sampling probe into the exhaust flue...

Page 17: ...e time of the sensor is strongly linked to the reagent consumption present in the sensor As their consumption increases features of the sensor are getting worse until their exhaustion after which it i...

Page 18: ...t the smoke probe from the analyzer Turn the lower part 2 of the water trap to the left to detach it from the upper part 1 Remove the part 3 by pulling it slightly Remove the used filter 4 and replace...

Page 19: ...975 info spain sauermanngroup com Sauermann Industrie ZA Bernard Moulinet 24700 Montpon France T 33 0 5 53 80 85 00 services sauermanngroup com Sauermann Australia 1 36 Campbell Avenue Cromer 2099 NS...

Reviews: