SATO SK-640PH Instruction Manual Download Page 14

23 

 

Backlight function 

 

This instrument is provided with a backlight function. Pressing the 

 

    key lights the LCD in green. 

This permits the pH readings on the LCD to be read easily even in a dark place. 
The backlight is automatically turned off approx. 20 seconds after it is turned on. This prevents the 
battery power from being consumed due to continuous use. 

 
 

About the AC adapter 

 

If the unit will be used for a long period of time, use of the AC adapter is recommended. 
Insert the AC adapter plug into the AC adapter jack on the unit. If an AC adapter other than the one 
provided by us is used, choose one that meets the specifications described below. 
 
  <Primary output> 
 

Voltage  

 

: 9VDC 

 

Current capacity 

: 500mA 

              Polarity   

 

:   

Diameter of jack   

: Ø 3.4mm (D

’ DIA Ø1.35) 

 
 

About the RS-232C output 

 

 

RS-232C output 
This instrument outputs data in real time on the pH, voltage and temperature while in the 
measurement mode or the MAX/MIN logging mode. 
When the unit is connected with a computer, data on the pH (or voltage) and the temperature 
can be imported into the computer for analyzing or saving. 
 

Note: The RS-232C cannot be used for output data while in the calibration mode or in the manual 

temperature setting mode. 

 

 

Connecting the cables 
Use the RS-232C communication cable dedicated to this product. 
(The cable is available as an option.) 

 

(1) Insert the plug of the optional RS-232C cable all the way into the RS-232C jack on the 

SK-620PH. 

 
(2) Next, connect the other side of the cable to the serial port on the computer. 

 
 

 

Communication standard and Communications protocol 
    Transmission speed 

: 2400bps 

    Data bit   

 

: 8 bits 

    Stop bit   

 

: 1 bit 

    Parity check 

 

: non 

 

    * Communications protocol 
 
    (1) pH display and the temperature (positive) display: 

 

 

 

 

 

Summary of Contents for SK-640PH

Page 1: ...Instruction manual for No 6422 00 Handy type pH meter Model SK 640PH No 3 Nishi Fukudacho Kanda Chiyodaku Tokyo 101 0037 Japan URL http www sksato co jp english N 11...

Page 2: ...e If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may increase Never disassemble or modify the unit Doing so may cau...

Page 3: ...mperature compensation Features Large LCD for easy reading The pH value and the temperature are indicated on a large LCD Auto Power Off The auto power off function turns off the power automatically if...

Page 4: ...e and MIN value The maximum and minimum voltage mV can also be logged To return to the measurement mode press the REC key for three seconds or more while in the MAX MIN logging mode Pressing this key...

Page 5: ...minimum pH or voltage value is displayed BAT Lights up when battery power is insufficient ATC Lights up when the temperature sensor is connected Display of time voltage Indicates the time elapsed sin...

Page 6: ...g the vinyl cover disconnect the sensor probe first and then remove in the reverse order of installation Note The vinyl cover is not waterproof splash proof according to the Japanese Industrial Standa...

Page 7: ...on the right 2 Installing the glass electrode protection bottle 2 1 Install the protection bottle to the sensor probe in the reverse order of removal 2 2 Hold the sensor probe upright so as not to spi...

Page 8: ...fter replacing the battery c When very accurate measurement is necessary For more accurate measurement calibration of about once a day is recommended For calibrating the SK 620PH prepare the following...

Page 9: ...alibration value is 4 04 7 Press the REC key to store the calibration value in the unit memory SA is displayed on the LCD to indicate that the calibration data has been stored in the unit memory Now t...

Page 10: ...ssible to perform calibration using the oxalate standard solution pH 1 68 at 25 C or borate standard solution pH 9 18 at 25 C CAL key REC key REC Key REC Key Measurement mode Calibration pH4 Calibrati...

Page 11: ...hich is capable of measuring both pH and temperature is used connect both the BNC plug pH and the diode plug temperature 2 Press the PWR key to turn on the unit The whole LCD lights up for approximate...

Page 12: ...21...

Page 13: ...eft unwashed the liquid measured tends to form a film on the surface of the electrode causing measurement errors Always wash the pH electrode every time after use Washing can be done with water distil...

Page 14: ...bout the RS 232C output RS 232C output This instrument outputs data in real time on the pH voltage and temperature while in the measurement mode or the MAX MIN logging mode When the unit is connected...

Page 15: ...d replacing battery The power to the unit is turned off The auto power off is turned on Cancel the auto power off on the main unit Before measuring Error code pH display section The sensor probe is no...

Page 16: ...an electric conductivity of approx 100 S cm or more The temperature of the sample largely fluctuated Make the sample temperature stable The pH values change with temperature Readings are abnormal The...

Page 17: ...unication conditions or protocols are not correct Check the communication conditions or protocols About the RS 232C output The serial port on the computer is not open Check that the serial port is ope...

Page 18: ...ent Temperature 0 to 50 Humidity lower than 80 rh no condensing Battery Life About 20 hours continuous Materials Case ABS resin Electrode Glass Probe Epoxy resin Dimensions Main unit W 72 H 182 D 31 5...

Page 19: ...28 6400 40 PH40 Carrying case 6400 50 PH50 Electrode protection bottle...

Reviews: