Warnings
Do not solder, short-circuit, disassemble, heat or fire a battery. Doing so may
cause an exothermic heat, a leakage of battery fluid or an explosion.
If the leaked fluid is spilt on your clothes or your skin, wash it off with water
and if it gets into your eyes, wash it off with clean water and seek a medical
attention immediately.
Battery poses a danger of suffocation or poisoning if swallowed. Keep
batteries out of reach of children.
When the battery is not in use, insulate the terminals with tape to prevent
contact with other metal objects or batteries. Otherwise, a short circuit may
occur and cause a fire or explosion.
For environmental conservation, dispose used batteries in compliance with
local rules and regulations.
Cautions
After measurement, put the sensing stem in the sheath attached.
At sheathing, make sure to confirm that the temperature is lower than 40°C.
Be sure not to continuously measure the object of high temperature for more
than 5 minutes. If so, the heat of the sensor conducts to the body and may
cause failure or the deform or breakage may occur
Do not bend or snap the sensor stem.
Do not use the unit in a water although you can operate it with wet hands
Do not clean the instrument with alcohol, thinner or any other solvents. Use a
clean cloth moistened with warm, neutral detergent water.
Never dissemble or alter the instrument.
Specifications
Name
Waterproof Digital Thermometer
Model
PC-9215
Ⅱ
Cat. No.
1747-50
Measuring Range
-30 to 150ºC
Accuracy
±1 ºC at 0 to 140ºC
±2 ºC at other range
Sampling Time
every 1 second
Power Requirement
1.5VDC (one LR44 button cell battery)
Battery Life
about 9 months (one hour use in a day)
Dimensions
Main unit: Ø47 × (t) 20 × (L) 176 mm (w/ senor)
Sensing section: Ø3.5 × (L) 116 mm
Materials
Main unit and sheath: ABS resin
Sensing stem: stainless steel
Waterproofness
Conform to JIS C0920 IPX5
Weight
about 25 g (inclusive of a battery)
Accessories
Sheath with clip (Ø10 x 144mm),
One LR44 button cell battery
* Heatproof temperature for meter and sheath is -10 to 40ºC
* The battery installed was used for monitoring purpose and its
battery life may be shortened.
* All specifications and appearances subject to change with or without
notice.
3-4, Kanda-kajicho, Chiyoda-ku, Tokyo 101-0045 Japan
URL: http://www.sksato.co.jp/english/
S.03