5-2. Key operation section 

 

 

 

 

 

  POWER key 

- Press this key to turn the main unit on. Press the key again to turn it off. 

  MODE key 

- Use this key to ensure the setting value.    In the HACCP mode, press this key to select mode. 

  HACCP key 

- Use this key to make a shift between to HACCP mode and measurement mode   

  MAX/MIN key   

- Use this key to display the maximum value and minimum value and clear the maximum value 

and minimum value.    In the setting mode, use this key to add set value 

  HOLD key   

- Press this key to hold (freeze) measured value. In the setting mode, press this key to subtract 

set value. 

 

 

6. How to Use 

6-1. Loading Batteries 

Set or replace the batteries with the following procedure.   

 

(1) Rotate the screw on the battery lid on the back of the 

main unit  counterclockwise  using a  blade  screwdriver 

or a coin until it comes free and then remove the lid. 

* Be careful not to be missing the rubber packing that is 

in the groove of the battery compartment. 

* If the body is wet, remove the battery lid after wiping off 

the moisture in order to prevent water invasion into the 

battery compartment. 

The screw does not come off the lid. Pulling it by force will cause breakage.

 

 

(2) Load the batteries so that each positive or negative terminal faces the correct direction as shown 

in the battery compartment. 

 

(3) Reinstall the battery lid after confirming that the rubber packing is properly in the groove and close 

the lid in the reverse order of the removal procedure. 

 

 

 

 

 

Summary of Contents for 8078-00

Page 1: ...Instruction Manual Waterproof Digital Thermometer Model SK 270WP No 8078 00 with a hole for wall mount type No 8078 20 without a hole for wall mount type S 01E...

Page 2: ......

Page 3: ...se this unit in water Do not use this unit in a place exposed to direct sunlight or near heating equipment Doing so may result in deformation of the casing or failure Do not use this unit in an enviro...

Page 4: ...tal thermometer to which a dedicated thermistor sensor probe can be connected Optional probes can be interchangeably connected with main unit 3 What in the box After unpacking please check the box con...

Page 5: ...probe it conforms to Protection level IP67 in JIS C 0920 and can be washed with running water NB This model is not suitable for use under water Impact resistance The outer protection rubber of the cas...

Page 6: ...red value is being held Low battery mark Blinks when battery power is insufficient MAX MIN Lights while the maximum minimum measured value is being displayed Lights the mark and set number in the HACC...

Page 7: ...r replace the batteries with the following procedure 1 Rotate the screw on the battery lid on the back of the main unit counterclockwise using a blade screwdriver or a coin until it comes free and the...

Page 8: ...he Thermometer The unit comes with a vinyl cover to keep it clean and to protect the unit from damage if dropped 1 Open the mouth of the vinyl cover and place the unit in it so that the side with a sl...

Page 9: ...section serial number of the unit and the calibration date are indicated Then measured value is indicated temperature measurement mode NB If the probe is disconnected when power is turned on error me...

Page 10: ...mount the unit on a wall No 8078 22 does not have the hole for wall mount After fitting the screw on a wall or a column mount the unit by using the hole for mount NB Shake the unit lightly to the left...

Page 11: ...IN key again MIN lights and to the lowest temperature value is indicated Press the MAX MIN key again MIN goes out and switch back to the real time display Under the temperature measurement mode or the...

Page 12: ...CP1 temperature setting The initial value for temperature C is set to 75 To change the preset temperature use the MAX MIN key to increase the value or the HOLD key to decrease the value After setting...

Page 13: ...d HACCP 3 Select the mode to be used Perform temperature measurement When the temperature reaches the set point the timer starts counting down When the time reaches the set point the LED turns on and...

Page 14: ...Measured value is lower than display range Internal processing is abnormal Remove batteries Then load batteries again after leave it for a while Replace the batteries to new ones A sensor signal is ab...

Page 15: ...0 Protection IP 67 rated when the probe is connected to the unit Operation ambient 10 to 60 C less than 80 rh no condensing Storage ambient 10 to 60 C no condensing Operational Functions HOLD MIN MAX...

Page 16: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english...