ERROR MESSAGE 

An error message appears on the LED. 

 

Display 

Cause  

Action 

Hi 

Temperature or humidity reading  is  beyond 

the upper display limit. 

Use only within the measuring range. 

Lo 

Temperature or humidity reading  is  beyond 
the lower display limit. 

Use only within the measuring range. 

Er 

Temperature display side  
  (while within the measuring range) 
Temperature sensor or sensor cord is faulty

 
The  sensor  probe  may  be  faulty. 
Contact  us  or  the  retailer  from  which 
the unit was purchased. 

Humidity display side  
 (while within the measuring range)    
Humidity sensor or sensor cord is faulty 

 
  Note: When "Er" is displayed in the temperature side, "Er" displays in also humidity display side.  
           When humidity value is out of the measuring range, "Er" is also displayed in the humidity 
           side.  Use only within the measuring range 
 
 
MAINTENANCE 
 

  Install the glass filter (For Type 1)  - Option -  

 
   Use the glass filter (option) in case circumstance for measurement  
   is not good enough due to dusts etc.  
 

  Remove the sensor cap and put the glass filter over the sensor 

section. 

② 

Set the rubber cushion on the glass filter as the illustration, and  

put the cap on. 

 

③ 

Maintenance of the sensor filter 

When the glass filter becomes dirty response will become slow and correct value will not be 
detected. Cleaning of filter at an interval of fixed period is recommended.  Wash the glass 
filter when it appears to be dirty.  After immersing the filter into the detergent for a while 
washes it with pure water.   After washing the filter dry it out. 
 

 

           CAUTIONS 

 . The filter is made of sintered glass.  Be careful in handling.  

. Response will be little slower when glass filter is attached. After washing the filter dry it out. 
  Shake the probe softly so that it can acclimate to the atmosphere. 

 
 

  Maintenance of filter (For Type 2) 

 
  When the filter becomes dirty, contact us or the retainer from which the unit was purchase.   
  The filter in Type 2 is built-in type. Do not replace it by yourself.

 

 

Summary of Contents for 8111-00

Page 1: ...INSTRUCTION MANUAL SK 110TRH Series Digital Thermohygrometers...

Page 2: ......

Page 3: ...r near heaters Doing so may result in failure or case will be deformed Do not use the unit in the circumstances where electric noises are generated Doing so may result in unstable display or large err...

Page 4: ...from which the unit was purchased FEATURES Measuring operation is simple by pressing the POWER ON OFF button Large LED font height 13 mm for easy reading Low battery mark blinks when the batteries ar...

Page 5: ...links at one second interval when the batteries are exhausted Replace the batteries with new ones Probes Type 1 Standard type General purpose monitoring environment in computer operation rooms factori...

Page 6: ...sor cap 9 Sensor filter built in 4 Humidity sensor HSK 10 Sensor cap L 5 Temperature sensor Thermistor 12 Sensor cap S 6 Sensing stem 13 Telescoping type stem HOW TO USE Loading batteries 1 To remove...

Page 7: ...dity values will be displayed being ready for measurement 2 Press POWER ON OFF button again for more than 1 second and the display will disappear and power will be off NB The probe and unit are unifie...

Page 8: ...on Use the glass filter option in case circumstance for measurement is not good enough due to dusts etc Remove the sensor cap and put the glass filter over the sensor section Set the rubber cushion on...

Page 9: ...ng condition Main body 0 to 50 C less than 80 rh no condensing Probe 10 to 60 C no condensing Sampling time About 1 sec Functions Low battery mark dual display for temperature and R humidity Power req...

Page 10: ...8 ACCURACY OF R HUMIDITY Humidity rh Calibration may not be accurate Temperature C...

Page 11: ......

Page 12: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan TEL 81 0 3 3254 8117 FAX 0 3 3254 8123 URL http www sksato co jp english R 01...

Reviews: