18
Parameter Settings: Overview
ο
= Factory setting;
√
= User-defined setting
Level 1
Level 2
Level 3
Level 4
Code
1.)
Setup
Bal.Scal
Ambient
V.Stable
Very stable conditions
1. 1. 1. 1
Weigh cell
conditions
ο
Stable
Stable conditions
1. 1. 1. 2
functions
(Filter adaptation)
Unstabl
Stable conditions
1. 1. 1. 3
V.Unstbl.
Very unstable conditions
1. 1. 1. 4
App.Filt.
ο
Final.Rd.
Final readout mode
1. 1. 2. 1
Application filter
Filling
Filling mode
1. 1. 2. 2
Stab.Rng.
1/4 dig.
(digit)
1. 1. 3. 1
Stability range
1/2 dig.
(digit)
1. 1. 3. 2
1 dig.
(digit)
1. 1. 3. 3
ο
2-dig.
(digit)
1. 1. 3. 4
4 dig.
(digit)
1. 1. 3. 5
8 dig.
(digit)
1. 1. 3. 6
Stab.
delay
No delay
1. 1. 4. 1
ο
Short delay
1. 1. 4. 2
Medium delay
1. 1. 4. 3
Long delay
1. 1. 4. 4
Taring
W/o Stb
W/o stability
1. 1. 5. 1
Taring
ο
W/ StAb
After stability
1. 1. 5. 2
Aut.zero
off
1. 1. 6. 1
Auto zero
ο
on
1. 1. 6. 2
Wt.Unit
For list of units,
1. 1. 7. 1
Basic
Gram to Newton
to
Weight unit
1. 1. 7.23
Display
ο
all
1. 1. 8. 1
Basic accuracy
Minus 1
One level lower
1. 1. 8. 2
Increment of the measured values one level lower
1. 1. 8. 3
Increment of the measured values two levels lower
1. 1. 8. 4
Increment of the measured values three levels lower
1. 1. 8. 5
Incrm. 1
Last digit single increment
1. 1. 8. 6
resolution by a factor of 10
1. 1. 8. 8
Cal./Adj.
ο
cal. ext.
Adjustment/calibration with factory-set weight
1. 1. 9. 1
Function of the
ο
e.cal.usr.
External calibration/adjustment
Q
key
with user-defined weights (factory-set on WZA25-NC)
1. 1. 9. 3
Cal.Int.
Internal calibration/adjustment only on models WZA…-NC
1. 1. 9. 4
lin.ext.
Linearization with factory-set weights
1. 1. 9. 6
lin.e.usr.
Linearization with user-def. weights
1. 1. 9. 7
str.prel.
Set preload
1. 1. 9. 8
clr.preload
Clear preload
1. 1. 9. 9
blocked
Q
Blocked
1. 1. 9. 11
Cal.routine
ο
Adj.
one sequence
1. 1. 10. 1
Cal./adj.
as required
1. 1. 10. 2
Cal.Unit
ο
Grams
1. 1. 11. 1
Weight unit
Kilogr.
Kilograms
1. 1. 11. 2
for calibration
Pounds
1. 1. 11. 3
Zero rng.
default.
(factory-set)
1. 1. 12. 1
Zero range
ο
2
Perc
.ent
1. 1. 12. 2
5
Perc
.ent
1. 1. 12. 3
10
Perc
.ent
1. 1. 12. 4
Int.zero
Zero at power-on default (factory-set)
1. 1. 13. 1
Power On
Initial zero 2%/max. cap
1. 1. 13. 2
ο
Initial zero 5%/max. cap
1. 1. 13. 3
Initial zero 10%/max. cap
1. 1. 13. 4
Initial zero 20%/max. cap
1. 1. 13. 5
Initial zero 50%/max. cap
1. 1. 13. 6
Initial zero 100%/max. cap
1. 1. 13. 7
on.tare
ο
On
1. 1. 14. 1
(Tare/Zero at
off
1. 1. 14. 2
Power/Zero-setting range)
Output
rate
ο
Normal
1. 1. 15. 1
Fast (five times faster)
1. 1. 15. 2