Safety Instructions

07.12.06

07.12.06

07.12.06

Date

Klausgrete

Klausgrete

Klausgrete

Name

Drawing no.

Title

Released by

Reviewed by

Written by

Revision

00

1.

The equipment may be used zone 2 and 22 hazardous areas. In zone 22, the device is approved for
use only in the presence of non-conductive dusts. The permissible ambient temperature is –10°C to
+40°C (14°F to 104°F). The equipment must be used indoors.

2.

The indicator may not be used as a portable device.

3.

Installation, operation, maintenance and repair of the equipment may only be performed by an
authorized technician and in compliance with applicable laws, rules and regulations, ordinances and
standards. In particular, make sure all work performed conforms to European Standards EN 60079-
14 (Electrical apparatus for use in potentially explosive gas atmospheres) and EN 50281-1-2
(Apparatus for use in the presence of combustible dusts). Installation, operation, maintenance and
repair of the equipment may only be performed when the equipment is completely disconnected from
the power supply.

4.

Make sure to follow the installation, operating, maintenance and servicing instructions given in the
manuals supplied.

5.

The cable glands must be fastened using a torque of 5 Nm. Use only 3 Nm to fasten the power
cable. The external connecting cables must be installed in a protective tube and secured to prevent
damage and stress caused by strain. The cable connections in the hazardous area must be secured
to prevent them from working loose.

6.

All metal parts (housing, column, etc.) must be electrically connected to the equipotential bonding
conductor so that any electrostatic charges can be conducted away from the equipment. The
equipment operator is obligated to connect a lead with a gauge of at least 4 mm² (cross section) to
the ground terminal (marked by the grounding symbol) located at the rear of the housing. The low
resistance of this connection to the grounding conductor busbar must be checked when the system is
installed at the intended place of use. Make sure the grounding conductor cannot be disconnected by
pulling on it. The shielding of the connecting cables may only be used for grounding when no
impermissible difference in voltage is generated and the shielding is able to conduct the potentially
occurring equipotential current.

Both the display and control unit and the weighing platform

must be connected to an equipotential bonding conductor

if no metallic connection (such as a 

column) is used. The standard display and control unit has no grounding terminal and thus must be
grounded by internal wiring.

7.

Before opening the equipment, completely turn off the power or make sure that no potentially
explosive atmosphere is present in the area or any other explosion hazard in the surrounding
area. Never connect or disconnect cables while the power is on in a hazardous area!

8.

Use equipment only in the temperature ranges indicated. Avoid exposing the equipment to
impermissible sources of heat or cold. Avoid heat build-up. Make sure the equipment is properly
ventilated.

9.

Ensure proper installation to avoid vagrant current (e.g., magnetic fields).

10.

Install the indicator (or the display and control unit supplied with the complete scale) in such a way
that mechanical hazard to the IP protection is kept to a minimum.

11.

Make sure that the equipment is installed and operated without being exposed to impermissible UV 
radiation (avoid direct exposure to sunlight).

12.

Ensure that the voltage rating or power supply is identical to your local power rating. When
connecting the equipment to the power supply, the laws valid in your country must be observed. Be
sure that the correct power supply cable is used.

13.

Prior to initial operation, make sure that there is no potentially explosive atmosphere or any other
explosion hazard in the surrounding area.

These safety instructions apply to the installation, operation, maintenance 
and repair of the equipment

Summary of Contents for Signum: Option Y2 SIWA

Page 1: ...Installation Instructions and Safety Information Sartorius Signum Option Y2 Models SIWR SIWA SIWS Complete Scales for Use in Zone 2 and 22 Hazardous Areas 98648 016 49...

Page 2: ...models are suitable for use as category II 3 GD electrical apparatus in hazardous areas designated zone 2 gases or 22 non conductive dusts Note Read these installation instructions and safety informa...

Page 3: ...solely responsible for any damage or injuries that occur when using cables not supplied by Sartorius General Provisions for Installing Signum Models Equipped with Option Y2 Signum models equipped with...

Page 4: ...ting the scale to power Observe all warnings and safety precautions Using the Equipment in Legal Metrology If the scale is to be verified make sure to observe the applicable regulations regarding veri...

Page 5: ...pare and install the cable as follows Guide the cable through the cable gland Close and tighten the cable gland in accordance with the applicable standards Remove the casing from a section of the cabl...

Page 6: ...Assembly Toothed lock washer cable toothed lock washer Screw or nut with locknut Additionally secure the cable with a clip strain relief For SIWS BBP models Ground the indicator Connect the cable to t...

Page 7: ...is any indication that safe operation is no longer warranted for example If there is visible damage to the connecting cable If the device no longer functions properly If the equipment has been stored...

Page 8: ...icians on location or our Service Center in Goettingen Germany Sartorius AG Service Center Weender Landstrasse 94 108 37075 Goettingen Germany Prior to disposal and or scrapping of the equipment any b...

Page 9: ...9...

Page 10: ...10...

Page 11: ...11...

Page 12: ...pplied e g printer No power supplied from connected device to scale These devices can be installed in zone 2 or 22 if they are suitable for category 3 in accordance with ATEX Directives Gas group IIC...

Page 13: ...Y2 or other weighing platform s suitable for zone 2 or 22 category 3G or 3D and that are permitted to be powered by nominal bridge supply voltages of at least 9VDC A D converter output with bridge su...

Page 14: ...of the housing The low resistance of this connection to the grounding conductor busbar must be checked when the system is installed at the intended place of use Make sure the grounding conductor cann...

Page 15: ...re they do not exceed the maximum surface temperature The outputs must have Ex nA electrical circuits 20 Chemicals and other agents that can corrode or otherwise damage attack the housing seals and ca...

Page 16: ...G Goettingen Germany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG The status of the i...

Reviews: