7
Startup
Install components
(6 - 4)
in the weighing chamber.
The balance has been factory-set to 220 volts.
How to change the voltage setting:
Adjust the voltage selector
(1)
to your local line voltage rating.
Write the new voltage setting on the manufacturer's label lo-
cated on the rear panel of the balance. Whenever you change
the operating voltage to 100/120 V, make sure to exchange the
installed T 80 mA fuse
(1)
for the T 160 mA fuse that comes
with the balance.
The fuse
(1)
is plugged into the voltage selector located on the
rear panel of the balance.
100/120 V-T 160mA
220/240 V-T 80mA
At the point of use, plug the power cable into a properly in-
stalled wall outlet. lf you use a wall outlet that does not have a
protective grounding conductor, make sure to ground the bal-
ance.
Level the balance using the leveling feet
(7)
so that
the air bubble is centered within circle of the level
indicator
(8)
.