38
Operating Instructions PMA 35001-X
Verification of Intrinsic Safety
Page
5
of
4
36953-761-60
Verification of Intrinsic Safety
2013-05-02
Dr. D. Klausgrete
Revision 00
PMA35001-X..
RS485 Data Interface
YDI05-Z.. interface converter
II (2) GD [EEx ib] IIC or
Z966 Zener barrier
4)
in YDI02-Z..: II (2) G [EEx ib] IIC
or
YCO01-Y interface converter
II (2) GD [EEx ib] IIC
or
II 3 (2)GD EEx nR[ib]IIC T4
Zone 1
Gas: IIC T4
Hazardous Area
Non-Hazardous Area
Z966
1)
pins A/J/K/N and M
Uo
12 V*
24V**
Io
82 mA /164mA
6)
Po
0.24 W / 0.48W
6)
Co
1.41 µF*
125nF**
Lo
5.52 mH
1.38mH
6)
Lo/Ro 147 µH/
57 µH/
6)
YDI05-Z..
2) 3)
Uo
12.4 V*
Ui
12.6 V*
24.8V**
25.2
V**
Io
130 mA***
Ii
any
Po
400 mW
Pi
any
Co
1.24 µF*
Ci
0
112nF**
Lo
0.4 mH*
Li
0
0.4 mH**
Lo/Ro 44 µH/ * / 22 µH/ **
YCO01-Y
2)
Uo
7.2 V*
Ui
12.6 V*
8.0 V**
25.2 V**
Io
207 mA***
Ii
1.5 A
Po
330 mW
Pi
2.5 W
Co
13.5 µF*
Ci
1 nF
8.4 µF**
Lo
0.7 mH*
Li
1.6 µH
0.7 mH**
Lo/Ro 36 µH/ * / 36µH/ **
Up to 7 additional
CIXS3
or
SIWS… / ISXS...
3)
Data cable:
Recommended: Sartorius cable
YCC485-X
with approx.
10µH/ohm and 120pF/m (wire/wire)
up to 1000m flexibly installed
.
RS485 data interface
5) 2)
(
LV4
)
Uo
7.2 V
3)
Ui 12.6 V
8.2 V
4)
Io
168 mA*
Ii
1.5 A*
Po
0.25 W
Pi
2.5 W
Co
13 µF
3)
Ci 300 nF
3)
Co
7.6 µF
4)
Ci 100 nF
4)
Lo
0.8 mH
Li
0 mH
Lo/Ro 118 µH/ohm
RS485 data interface
5) 2)
(
UNICOM
.
LV2
)
Uo
7.2 V
Ui 12.6 V
Io
127 mA*
Ii
1.5 A*
Po
0.273 W
Pi
2.5 W
Co
11.3 µF
Ci 0
Lo
2 mH
Li
2 µH
Lo/Ro 118 µH/ohm
RS485 data interface
2)
Ui see below
Uo
5.2 V
Ii
see bwlow
Io
210 mA***
Pi any
Po
263 mW
Ci 260 nF
Co
60 µF
Li
0 mH
Lo
600 µH
Lo/Ro
125 µH/
Ui
±
12.4V
12.0V
7.2V
Ii
130 mA***
164mA***
any
Note:
1: Per circuit
2: Combined circuits
3: Only two RS232 connections are used on the YDI05-Z
4: BAS01ATEX7005; II (1) GD [EEx ia] IIC
5: Data for CIXS3
6: Both channels connected on Z966
*: Versus GND; **: Between the lines; ***: Resistively limited
PMA35001-X..
II 2 G Ex ia IIC T4 Gb
Summary of Contents for PMA.Power Series
Page 29: ...Operating Instructions PMA 35001 X 29 EC Declaration of Conformity...
Page 30: ...30 Operating Instructions PMA 35001 X ECDeclaration of Conformity...
Page 32: ...32 Operating Instructions PMA 35001 X EC Verification...
Page 33: ...Operating Instructions PMA 35001 X 33 EC Verification...
Page 34: ...34 Operating Instructions PMA 35001 X EC Verification...
Page 40: ...40 Operating Instructions PMA 35001 X Certificate of Compliance...
Page 41: ...Operating Instructions PMA 35001 X 41 Certificate of Compliance...
Page 42: ...42 Operating Instructions PMA 35001 X Certificate of Compliance...
Page 43: ...Operating Instructions PMA 35001 X 43 Certificate of Compliance...