24
Miras 2
Miras 2
Special Error-Dependent Codes:
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
E R R
n n1 n2
CR LF
n – n3 contains an error code of up to 3 digits.
Data Input Formats
You can enter certain commands to control weighing platform functions through the SBI
interface.
As with data output, data is input as 7-bit ASCII characters; hardware and protocol con-
figuration are identical to those for data output.
Formats:
ESC K CR LF
ESC K K1 – CR LF
ESC : Escape
K : Command character
K1 : 2
nd
command character (number)
-
: Underline
CR : Carriage return
LF : Line feed
24