Malfunctions
8 Malfunctions
Problem
Cause
Remedy
No segments appear
on the weight display
− No AC power is available
− Check power supply
The weight readout
shows “
LOW
”
− No weighing pan on the
scale
− Position the weighing
pan
The weight readout
shows “
HIGH
”
− Weighing capacity
exceeded
− Unload the scale
The weight readout
changes constantly
− Unstable ambient
conditions
− Excessive vibration or
draft
− Weighing pan is being
affected at some point
by outside influences
− Set up scale in
another area
− Adjust the scale
settings (see Chap-
ter “3.5 Menu List,”
page 13)
The weight readout is
obviously incorrect
− The sample is not stable
− Scale not tared before
weighing
− Tare before weighing
No weight value is shown
and the padlock symbol
H
is displayed
− PC connection to the
scale has been inter-
rupted, activating the
“LOCK” function
− Manual “LOCK”
function is active
− Access the scale
operating menu to
make the necessary
adjustment and switch
off the “LOCK”
function
− Check the connection
− Switch off the manual
“LOCK” function
Summary of Contents for EVO1X
Page 2: ......
Page 43: ...Operating Manual EVO1X LAB1X 43 Conformity Licenses...
Page 44: ...44 Operating Manual EVO1X LAB1X Conformity Licenses...
Page 67: ...Operating Manual EVO1X LAB1X 67 Conformity Licenses...
Page 68: ...68 Operating Manual EVO1X LAB1X Conformity Licenses...
Page 69: ...Operating Manual EVO1X LAB1X 69 Conformity Licenses...
Page 70: ...70 Operating Manual EVO1X LAB1X Conformity Licenses...