The following data are stored permanently in DocuClip
®
:
– Initial calibration including date and time
– pH and voltage in mV
– Nominal values of the buffers used
– Serial number of Docu-pH
Meter
– Serial number of the electrode
– Temperature
§
View data of installed electrode/DocuClip
®
:
See Setup-Menu: Calibration functions: pH/mV: Data DocuClip
®
Enter serial number of the Electrode on Docu-pH+:
Step Press
key
Display
1. Connect DocuClip
®
2. Start “Serial number input”
Start
softkey
3. Set “Electrode number” (S/N)
Repeatedly
press
Q
softkey,
O
softkey
4. Confirm “Electrode number”
OK
softkey
!
Initial calibration: Please work
with particular care. The initial
calibration documents the degree
of effectiveness of the new electrode.
This calibration is the reference point
for the later analysis whenever there
is a change in performance.
27
Initial Calibration
Input
Serial No.
Cancel
Initial Calibration
Input
Electrode No.
Calibration
Buffer 1
Bufferset:
Cancel