Installation and Operation Manual

arium 611VF

Water Purification System

85030-513-68

Summary of Contents for arium 611VF

Page 1: ...Installation and Operation Manual arium 611VF Water Purification System 85030 513 68...

Page 2: ...tive Mode 23 Maintenance and Servicing 24 Sanitization 24 Cartridge Replacement 25 Ultrafilter Replacement 26 Replacing the Final Filter 26 Cell Cleaning 27 UV Bulb Replacement 28 Activate the UV Time...

Page 3: ...emove the plug from the electrical outlet prior to maintenance and servicing of arium 611VF 3 Safety Information Warning Danger of fire or explosion Do not use in the presence of flammable or combusti...

Page 4: ...erated by trained personnel Operate arium 611VF only using original accessories or replacement parts If you modify this water purification system on your own without consulting Sartorius the performan...

Page 5: ...played on a digital display Item No Description 1 Left and right door of arium 611VF 2 Product water outlet with 0 2 micron filter and bell assembly 3 Display and control panel 4 Recessed grip for ope...

Page 6: ...ms of the different menus The Y key moves the cursor one line up the y key one line down ENTER Key Use this key to select a menu item you have highlighted with the cursor Display During Normal Operati...

Page 7: ...Control panel and display of arium 611VF Flow chart of arium 611VF 7...

Page 8: ...de and screw to the assembly Route the water tubing through the 4 holes inside the door and connect to the color coded connectors Follow the course of dotted lines in the figure Positions 1 and 3 data...

Page 9: ...Mounting the upper display dispenser unit Mounting the lower display dispenser unit 9 1 2 4 3...

Page 10: ...from the electrical outlet O Open both doors of the unit Remove ultrafilter from packaging and remove protective covers from fittings Connect the tubing connectors to the ultrafilter module as shown i...

Page 11: ...asteners may result in injury to the operator and or damage to the equipment A wall bracket 2 is included in the shipment to enable you to attach the system securely to the wall so that it occupies a...

Page 12: ...ection of the 611 system Connect the free end of the tube to the appropriate water supply adapter Push the tube all the way into the adapter to seat it properly approx 20 mm Install the tubing adapter...

Page 13: ...upper ring of the spacers Push both safe guard slides all the way under the upper ring of the spacers as shown in the figure opposite Place the cartridge pack with the blue dot left into the housing a...

Page 14: ...he MM DD YY format and the time in the 12 hour format Choose a format and press ENTER to confirm Select Time Date and use the arrow keys Y or y to change current Time Date settings Press ENTER after e...

Page 15: ...l for the product water in MO cm or S cm When the water quality falls below this Set Point cartridge packs should be replaced The factory setting for the Set Point is 10 5 MO cm To select the Set poin...

Page 16: ...he instructions on the next few screens and press ENTER after each is complete Note Make sure system is depressurized before attempting to inject sanitization liquid Inject the sanitization liquid wit...

Page 17: ...Operation Mode Then open the dispensing valve and flush 5 liters of water to drain Close the dispensing valve LONG FLUSH 5 min Push ENTER to Start UF FLUSH Short Long MAINTENANCE UV Life Timer Saniti...

Page 18: ...and press ENTER to confirm With the Print Interval Option the system can be set up to print data at regular intervals If you have selected Print Interval you can use the control keys for the cursor t...

Page 19: ...le you have measured a 1 liter volume in 1 minute Your measured flow rate is 1l min If you need 10 liters to fill a container 10 l 1l per minute 10 minutes dispense time Note If you are using a filter...

Page 20: ...he red cartridge back in the system close the doors connect the system to the AC outlet and power on the arium water system I Press MENU to get into the Main Menu Use Y or y to select Utilities Press...

Page 21: ...n Danger of water spillage Press MENU to get into the Main Menu Use the cursor to select Mainte nance Press ENTER to confirm Use the DOWN arrow y to select Calibration Press ENTER to confirm After pre...

Page 22: ...ssembly of the final filter and Push the draw off lever to the right to dispense the product water After dispensing the water pull the draw off lever forward to close the valve Place the protective ca...

Page 23: ...packs replaced When the system is inactive the display shows If you need to shut down arium 611VF for an extended period of time you should completely drain it and remove all cartridge packs and the...

Page 24: ...urity levels To sanitize the arium 611VF Press MENU to get into the Main Menu Select Maintenance and press ENTER to confirm Select Sanitization and press ENTER to confirm Follow the instructions on th...

Page 25: ...e housing to the lower left side you will find a blue labeled connector on the right a red connector To facilitate connection of the adapter moisten O rings on the adapter for the cartridge pack Place...

Page 26: ...at the O ring on the blue adapter is seated in a vertical position before connecting the ultrafilter Push the ultrafilter module back into the holder so that it engages The connection fitting on the s...

Page 27: ...ble 2 under the yellow hood covering Remove the cell 1 from the black socket and unscrew it Carefully remove the O ring 3 before cleaning the cell Clean the cell in a mild detergent solution using a s...

Page 28: ...lug 1 from the black plastic cap 2 Unscrew the top plastic cover from the stainless steel chamber of the UV bulb see top figure on the next page Hold a beaker 300 500 ml size under the black plastic c...

Page 29: ...U to get into the Main Menu Select Maintenance and press ENTER to confirm Select UV Timer by using y and press ENTER to confirm Press ENTER to reset the timer The internal timer is reset and will remi...

Page 30: ...wer switch O and disconnect it from the AC power outlet The fuse drawer with two safety fuses is located on the left side of the housing above the main power switch Gently press the locking device of...

Page 31: ...y Resistivity cell dirty Clean cell see section on Cell Cleaning on page 30 The system is not calibrated Perform calibration see section on System Calibration on page 21 High endotoxin level The syste...

Page 32: ...inlet pressure of 7 bar Product water capabilities Quality of the purified water 18 2 MO cm at 25 C RNase DNase DNA free Bacteria 1 CFU 100 ml TOC 1 ppb Endotoxin 0 001 EU ml Flow rate Up to 1 5 l min...

Page 33: ...r low TOC level application in systems with an UV lamp chamber Will be used on the right red side of the system 5441307H4 NO Sartopore 2 150 Final Filter 2 611CKDU Consumable kit with 611CPD for DI wa...

Page 34: ......

Page 35: ......

Page 36: ...torius AG Goettingen Germany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG The status...