5
Duty button
When the ON/OFF switch (4) is set to "on", the balance can also be switched off and back on again with the
duty button (6).
lf there is no weight change for at least 0.5 minutes, the display will read "---" followed by switch-off.
To switch on: depress button (6) - display will show multiple eight-check-
To switch off:
depress button (6) for approx. 3 secs., after which display will read "---" confirming balance is switching off.
Accessories
Calibration weight
(1x500 g)
cat. no. 707216
Carrying case
cat. no. 606707
NC battery*
cat. no. 629809
The manufacturer's guaranty does not cover balance damage resulting from battery / NC battery leakage.
Data lnput with 4 function keys cat. no. 609001
* Balance is suitable for use with commercially sold NC batteries (e.g. 6x501 RS Mignon) or batteries
(e.g. 6x1.5 V Mignon cells).
Warning!
Use only NC batteries with a minimum permanent Charge current of 50 mA.
Disregard of this specification may result in leakage.
Technical Data
Model
1002 MP9
1006 MP9
Weighing range
505 g
5050 g
Readability
0,1 g
1 g
Standard deviation (reproducibility)
≤
±0,1 g
≤
±1 g
Max. deviation from linearity
≤
±0,1 g
≤
±1 g
Taring range (by subtraction)
505 g
5050 g
Stabilization time (typical)
~ 2s
Integration time
400/800 ms
Display update
≥
200 ms
Ambient temperature range
273 K ... 313 K (0
o
C ...+40
o
C)
Sensitivity drift (ambient temperature
range 283 K … 303 K)
≤
±7
∙
10
-5
/K
≤
±4
∙
10
-5
/K
Power supply
from power supply component 220 V/110 V
Consumption 0,35
W
Batteries
6x 1,5 V Mignon cells
Service life
approx. 30 hrs. in continuous operation
NC batteries
six 501 RS Mignon
Service life
approx. 18 hrs. in continuous operation