Installation 9

Removing the Spacers (As Needed)

If you want to install cassettes with a different filtration area in the filter holder, 
you must remove any spacers mounted. To do so, complete the following steps:

t

t

Remove the upper tensioning nut (3), the PEEK washer (2) and the end bolt (1b) 
from the top pull rod.

t

t

Remove the desired number spacers (1a) and save them.

Tensioning the Cassette

Danger due to improper installation of the pull rod!

When installing the pull rod make sure that the spacer, terminal spacer and PEEK 
plain washer are assembled correctly. The wrong configuration can lead to incorrect 
tightening torques and to destruction of the filter holder. 

 

− The PEEK plain washers must always be installed on the terminal spacer; the flat 

side of the terminal spacer is especially designed to accommodate the PEEK plain 
washers.

Before each use, the PEEK plain washers must be cleaned with water in order  

to completely maintain the functionality of the filter holder. 

 

− After every 10 uses, the plain washers should be replaced with new ones.

t

t

Push the spacer (1a, double conical) and the terminal spacer (1b, simple taper) 
onto the top pull rod in such a way that the taper of the spacers fits into the taper 
of the clamping plate (4).

t

t

Check to make sure the flat site of the terminal spacer (1b) is facing the outside.

t

t

Clean the PEEK plain washer (2) with water and insert it onto the pull rod.

t

t

Screw the tensioning nut (3) loosely onto the top pull rod.

t

t

From above, place the top pull rod in the groove of the product plate, cassette and 
clamping plate.

t

t

Check that the cassette is seated correctly and tighten upper and lower tensioning 
nuts by hand (5).

t

t

Thread the socket head onto the torque wrench.

t

t

Set the correct tightening torque on the torque wrench.

t

t

Uniformly tighten both tensioning nuts with the torque wrench by alternating 
between up and down.

For information on connecting the cassette to the crossflow filtration units, please 
refer to the cassette installation instructions.

1a

1b

2

3

1a

1b

2

3

5

4

Summary of Contents for Sartocon Slice SU

Page 1: ...Installation Instruction Installationsanleitung Sartocon Slice SU Filter holder Filterhalter 85037 545 73 85037 545 73 Vers 07 2014...

Page 2: ...English page 3 Deutsch Seite 15...

Page 3: ...Use and Foreseeable Misuse 6 2 4 Danger due to Improper Handling 6 3 Device Description 7 4 Installation 8 4 1 Unpacking 8 4 2 Installation Location 8 4 3 Inserting the Cassette 8 4 4 Removing Casset...

Page 4: ...lost or misplaced please contact Sartorius for a replacement or download the latest version from our website www sartorius com In addition to the installation instructions all generally valid legal a...

Page 5: ...lt if it is not avoided This symbol denotes a possible danger with risk that moderate or minor injury may result if it is not avoided This symbol denotes a danger with low risk that could result in ma...

Page 6: ...on with Crossflow Sartocon Slice Self Contained cassettes Any other use beyond this is considered improper use Observe the information in Chapter 7 Technical Data page 13 Danger due to use of unsuitab...

Page 7: ...of application include the concentration and diafiltration of antibodies vaccines and IgG for example When cassettes with different installation height are used the corresponding number of spacers are...

Page 8: ...wing tightening torques are recommended for devices with bronze tensioning nuts trapezoidal threads and PEEK plain washers July 2014 Please note that this information is only given as general recommen...

Page 9: ...mpletely maintain the functionality of the filter holder After every 10 uses the plain washers should be replaced with new ones t t Push the spacer 1a double conical and the terminal spacer 1b simple...

Page 10: ...formly and alternatingly unscrew the upper and lower tensioning nuts 1 2 t t Lift the top pull rod 3 upwards t t Remove the cassettes from the filter holder t t Dispose of them together with the accom...

Page 11: ...the cleaning agents used are compliant materials stainless steel 316L Danger of injury due to improper handling Observe the safety instructions for the cleaning agents The use and disposal of cleaning...

Page 12: ...r dirt the tensioning forces will be affected and the cassettes may be destroyed Fault Possible causes Corrective measures Filter holder does not seal The clamping pressure is insufficient Set the tor...

Page 13: ...ber of cassettes 1 7 3 Spare Parts Order Number Part Description BB 8775004 Sartocon Slice single use filter holder 17521 022 Socket head SW17 17521 023 Torque wrench 10 60 Nm 1ZLI V0001 Spacer Sartoc...

Page 14: ...ugust Spindler Strasse 11 D 37079 Goettingen Phone 49 551 308 0 Fax 49 551 308 3289 In countries that are not members of the European Economic Area EEA or where no Sartorius subsidiaries or dealership...

Page 15: ...sgem e Verwendung vorhersehbare Fehlanwendung 18 2 4 Gefahr durch unsachgem e Handhabung 18 3 Ger tebeschreibung 19 4 Installation 20 4 1 Auspacken 20 4 2 Aufstellort 20 4 3 Cassette einsetzen 20 4 4...

Page 16: ...der Installationsanleitung k nnen Sie Ersatz anfordern oder die aktuelle Anleitung von der Sartorius Website herunterladen www sartorius com Erg nzend zur Installationsanleitung sind allgemeing ltige...

Page 17: ...iko die eine mittelschwere oder leichte K rperverletzung zur Folge haben kann wenn sie nicht vermieden wird Dieser Hinweis kennzeichnet eine Gef hrdung mit geringem Risiko die Sachsch den zur Folge ha...

Page 18: ...Crossflow Cassetten Sartocon Slice Self Contained vorgesehen Jeder dar ber hinausgehende Gebrauch gilt als nicht bestimmungsgem Beachten Sie die Angaben im Kapitel 7 Technische Daten Seite 25 Gefahr...

Page 19: ...tration Die Einsatzbereiche umfassen z B die Konzentrierung und Diafiltration von Antik rpern Vaccinen und IgG Beim Einsatz von Cassetten anderer Bauh he wird eine entsprechende Menge Distanzbuchsen e...

Page 20: ...t Bronze Spannmutter Trapezgewinde und PEEK Unterlegscheibe Juli 2014 werden folgende Anzugsdrehmomente empfohlen Bitte beachten Sie dass es sich bei diesen Angaben nur um allgemeine Empfehlungen f r...

Page 21: ...die Funktionalit t des Filterhalters vollst ndig zu erhalten Nach jedem zehnten Gebrauch sollten die Unterlegscheiben durch neue ersetzt werden t t Schieben Sie die Distanzbuchse 1a doppelt konisch g...

Page 22: ...t t L sen Sie gleichm ig und abwechselnd die obere und untere Spannmutter 1 2 t t Heben Sie die obere Zugstange 3 nach oben ab t t Nehmen Sie die Cassetten aus dem Filterhalter t t Entsorgen Sie dies...

Page 23: ...Stellen Sie sicher dass die eingesetzten Reinigungsmittel materialkonform sind Edelstahl 316L Verletzungsgefahr durch unsachgem e Handhabung Beachten Sie die Sicherheitsvorschriften zu den Reinigungsm...

Page 24: ...winde f hren zu falschen Anzugskr ften die die Cassetten zerst ren k nnen St rung M gliche Ursachen Abhilfema nahmen Filterhalter undicht Der Anpressdruck ist nicht ausreichend Stellen Sie den Drehmom...

Page 25: ...llnummer Teilebeschreibung BB 8775004 Sartocon Slice Single use Filterhalter 17521 022 Stecknuss SW17 17521 023 Drehmomentschl ssel 10 60 Nm 1ZLI V0001 Distanzbuchse Sartocon Slice 2 Stck doppelt koni...

Page 26: ...dim Biotech GmbH August Spindler Strasse 11 D 37079 G ttingen Telefon 49 551 308 0 Fax 49 551 308 3289 In L ndern die keine Mitglieder des Europ ischen Wirtschaftsraumes sind oder in denen es keine Sa...

Page 27: ......

Page 28: ...ghts reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius Stedim Biotech GmbH The status of the information...

Reviews: