power  from  its  output.  To force another recharging the connector to the Myriam has to be 
removed and inserted again.

Filter replacement

Important  note:  Use  only filters of the specified type. The filter material affects directly the 
quality of the alpha spectroscopy. Wrong filter material causes malfunction.
To  replace  the  filter,  loose  the  small  screw  on  the  back  of  the  pivot  mounted  arm.  Swivel 

away the arm towards the 
front  panel  for  free  filter 
access.  Unscrew  the  filter 
nut 

anti-clockwise 

and 

remove  the  old  filter.  Do 
not  touch  the  surface  of 
the 

semiconductor 

detector! Start a sample to 
make  the  pump  running. 
The  new  filter  can  be 

placed now very easily because the pressure of the pump fixes the filter on the air inlet. Take 
care for the filter material because it is very sensitive against bending. Do not use any sharp-
edged tool. Fix the filter nut slightly using two fingers. Too strong fixing may damage the filter 
as well.

Maintenance

The Myriam contains a high quality membrane pump to collect the aerosols on the filter. This 
ensures  a  long  life  time  and  a  very  stable  flow  rate  during  the  operation.  Like  all  moved 
mechanical  part  also  the  pump  still  undergoes  an  abrasion  process.  Therefore  the  proper 
function should be checked periodically. If the Myiam is permanently used, the check can be 
carried out monthly after the filter replacement. It is sufficient to do that once per year if the 
instrument is only used from time to time (e.g. during calibration). 

If the flow rate drops below 10% of the value stated in the calibration certificate, the pump 
must be replaced because the flow rate affects the sensitivity (linear relationship). 
A  calorimetric  mass  flow  sensor  should be used for the flow measurement because those 
types offer a very low flow resistance. Insert a new filter for the measurement.

To determine the flow rate, the detector head must be pivoted. Then a special adapter with 
elastomere sealing (available from SARAD) can be pressed tightly at the filter nut to suck the 
air through the air flow sensor. 

Please check the Myriam for loose parts (detector head, pivot arm, filter nut).

Summary of Contents for MyRIAM

Page 1: ...Myriam_EN_10 04 08 doc c SARAD GmbH User Manual MyRIAM Version 04 2008 SARAD GmbH Tel 49 0 351 6580712 Wiesbadener Stra e 10 FAX 49 0 351 6580718 D 01159 Dresden e mail support sarad de GERMANY intern...

Page 2: ...ons 6 Gamma interference high count rates and shock rejection 6 Data transfer 7 Battery and Charger 7 Filter replacement 8 Maintenance 8 Technical Data 9 Detection thresholds and detection limits 10 S...

Page 3: ...olled area Stopping the measurement and reading saving the acquired data Recharging the internal batteries if necessary At the end of an observation period Starting the measurement using the Filter An...

Page 4: ...ffers a special operation mode where the pump remains switched off during the measurement The influence of the Radon daughters will become negligible and the detection limit may decreased as required...

Page 5: ...he button for at least five Seconds If the Myriam is turned on again the data can be read via the infrared interface Pressing the button for a second time will start a new sample and the old data are...

Page 6: ...he calculated dose value drops below the adjusted limit This situation can take place if the alert has been initiated by a high Gamma radiation or by a strong mechanical shock or continuous mechanical...

Page 7: ...the measurement must be stopped before Battery and Charger The Myriam is powered by internal NiMH batteries Using the Sampling Mode running pump an operation period of more than 20 hours is guaranteed...

Page 8: ...as well Maintenance The Myriam contains a high quality membrane pump to collect the aerosols on the filter This ensures a long life time and a very stable flow rate during the operation Like all move...

Page 9: ...Integration Interval 1 255 Minutes one Minute steps Detector 150 mm ion implanted silicon detector Pump 0 25 L min Filter 0 8 m PTFE operable 1 Month at normal dust exposure Alpha Spectroscopy 2 8 10...

Page 10: ...limits are given for the most radio toxic nuclides Pu 239 and Sr 90 without Y 90 The dose coefficients e50 of those nuclides are derived from the ICRP68 standard for inhalation 4 7E 5 Sv Bq Pu 239 an...

Page 11: ...ion interval are stated below Fig 2 Detection Limit Threshold 3 for Ti 60 min After finishing the observation period e g one month the inhalation dose for the whole period can be determined by using t...

Page 12: ...he first time the Windows system detects the new hardware and starts the installation assistant The installation assistant will ask for a sufficient driver which is available on the Myriam installatio...

Page 13: ...le can be started also by pressing the button STOP Stops the running sample if not yet stopped and loads the acquired data and the instrument information from the Myriam to the PC During the data tran...

Page 14: ...p or using the Filter Analysis mode without pump Using the radio button group Push Button locks and unlocks the push button of the Myriam The edit fields in the lower part are important for the dose c...

Page 15: ...asurement After completion of each sequential interval a data record will be stored into the internal data logger If the check box Beep is marked the internal buzzer will sound for each keystroke Sets...

Page 16: ...the set up dialog and the graphic window are the same like selected in the main window before independent on the recent settings of the instrument If the parameters have been transmitted to the instr...

Page 17: ...specially at low exposures and short integration intervals the true value may be lower or higher than the displayed one It is strictly recommended to take a look at the acquired time distribution If n...

Page 18: ...MyRIAM 18 Manual_Myriam_EN_10 04 08 doc c SARAD GmbH...

Reviews: