7
3.5
High Frequency CMRR adjustment
a.
Referring to Fig. 11, connect output BNC to the oscilloscope has good grounding.
b.
Connect both input lines to a 155 Vp-p, 20 kHz, sine wave signal source.
c.
Put SW1 at 1/100.
d.
Set VOLTS/DIV switch on oscilloscope’s panel to 1mV and the TIME/DIV switch to 0.1ms.
e.
Adjust VC1 slightly to make the amplitude of displayed waveform as small as possible.
Finally the amplitude should be less than 4.9 mVp-p (after excluding noise)..
Fig. 11
3.6
Repeat 3.2 and 3.5 adjustment until both adjustments are O.K.