Sapcon ELIXIR-T-UNI Instruction Manual Download Page 5

User Manual & Datasheet V 2.1

1

Introduction

Elixir-T-Uni is a Vibrating Fork Level Limit Switch for

liquids. It is suitable for level detection in storage tanks,
mixing container and pipelines, for all kinds of liquid which
do not react with SS 316, viscosity does not exceed
10,000 cP and temperature lies within

0

C to

80

C. It is

suitable for most of the applications where float switches
were previously employed, as well as in places where
float switches were not appropriate (due to deposition,
turbulence, stresses and air bubbles). Elixir-T-Uni ignores
foam deposition on its fork tines.

Figure 1: Elixir-T-Uni

2

Operating Principle

Elixir-T-Uni works on the principle of continuous vibration

of the fork at its natural frequency.

A specially shaped

Tuning Fork is kept vibrating by piezo electric element.
The resonance frequency changes when the tines are
immersed in any fluid.The change in frequency is detected
and used is for switching.

Figure 2: Part Diagram

3

System Description

Elixir-T-Uni vibrating fork level switch detects the res-

onant frequency change and changes the output signal
according to fail-safe mode selection. Typically, the output
is fed to a PLC which then uses it to control the process.

4

Features

Power Supply: 18-30 V DC

Output: Open collector PNP, LED Indication

Operational temperature:

0

C to

150

C (Fork)

Hygienic finish with 0.8 Ra available

No calibration required at site, easy to install

Compact in size

Service friendly plug-in connection

Rugged stainless steel housing

Operational safety ensured

Immune to foam build-up

External circuit testing by test magnet

5

Applications

Elixir-T-Uni is suitable for the following applications and

industries:

Chemicals

Pesticides

Edible Oil

Breweries

Food Industry

Dairy Industry

Packaging Industry

Pharmaceutical
Industry

6

Dimensional Layout

Dimensional drawing for a fork length of 44 mm is shown

in Figure 3

193

143

68

50

44

32

For Hygienic Application:
Surface Finish in Ra = 0.8µm

Figure 3: Dimensional Diagram

(All dimensions in mm)

Sapcon Instruments Pvt.Ltd.

R

5

Summary of Contents for ELIXIR-T-UNI

Page 1: ...ELIXIR T UNI Compact Vibrating Fork INSTRUCTION MANUAL Version 2 1 ...

Page 2: ...al Specifications 7 10 Installation Guidelines 8 10 1 Procedure for Installation 8 10 2 Electrical Connections 8 10 2 1 For Fail safe High 8 10 2 2 For Fail safe Low 8 10 3 Output Check Test 9 10 4 Output Indications 9 10 5 Error Indications 9 11 Order Code 10 12 Customer Support 11 List of Figures 1 Elixir T Uni 5 2 Part Diagram 5 3 Dimensional Diagram All dimensions in mm 5 4 Fail safe High 8 5 ...

Page 3: ...User Manual Datasheet V 2 1 List of Tables 1 Electrical Specifications 6 2 Application Specifications 6 3 Mechanical Specifications 7 4 Error Indications 9 Sapcon Instruments Pvt Ltd R 3 ...

Page 4: ...l instrument housing in terms of dimensions color and design Please refer to GA drawings for dimensional details Values of performance described in this manual were obtained under ideal testing conditions Hence they may differ under industrial environment and settings General Instructions Instrument shouldn t block the material filling inlet Secure the cover of housing tightly Tighten the cable gl...

Page 5: ... used is for switching Figure 2 Part Diagram 3 System Description Elixir T Uni vibrating fork level switch detects the res onant frequency change and changes the output signal according to fail safe mode selection Typically the output is fed to a PLC which then uses it to control the process 4 Features Power Supply 18 30 V DC Output Open collector PNP LED Indication Operational temperature 0 C to ...

Page 6: ... mW 24 V DC 168 mW 12 V DC With Load 3W 24 V DC Table 1 Electrical Specifications 8 Application Specifications Please refer to Table 2 for Application Specifications PARAMETER VALUE Response Time Covered Uncovered Value 0 5 second 1 second Repeatability 0 5mm Switching Hysteresis 3mm 0 5mm Fork Resonance Frequency 1500Hz approx Minimum Liquid Density 0 7gm cc Maximum Viscosity of Liquid 10 000 cp ...

Page 7: ...lamp As per user specification Hygienic Flash Mounted Socket SS 316 with 1 BSP Electrical Connector Valve Connector Gland Type PG 11 PG 9 cable gland Dimension Refer to Figure 3 Active Fork Length 44 mm and 100 mm Probe Length 68 mm 105 mm 125 to 1000 mm For 44 mm Fork Length 126 mm For 100 mm Fork Length Process Temperature Up to 150 C Sensing Fork Material SS 316 Fork Finish Hygienic Fork Standa...

Page 8: ... Electrical Connections 10 2 1 For Fail safe High If a device is mounted at top of the tank then follow the instructions given below for Electrical Connections First of all connect the positive terminal to pin no 3 of the device Now connect the negative terminal to pin no 1 of the device Connect fuse between positive terminal and pin no 3 of the device Between pin no 1 and 2 the output can be take...

Page 9: ...ks 1 Second It is because of Serial Communication Stop Serial Communication Red LED Blinks 20 Seconds Oscillator Failed Fork is not connected Blue LED Blinks Slowly 1 Second Temperature goes above 100 C Blue LED Blinks Fast 300 MiliSec onds Temperature goes above 110 C Blue LED Glows Permanently Temperature goes above 120 C Table 4 Error Indications 2 Fork Stops Vibrating When Red LED blinks conti...

Page 10: ...xfcz Should you require further information regarding installation use or working of the instrument please don t hesitate to contact us Kindly provide the following information at the time of contacting Instrument Model and Serial Number Purchase Order Number and Date of Purchase Description of the query Your contact details In an attempt to serve you better we are open seven days a week 9 30am to...

Reviews: