User Manual & Datasheet V 1.2
ENTER
CASPER
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 19: Cover Delay Switch Position
ENTER
CASPER
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 20: Setting Cover Delay
•
Delay is entered, but not saved. To save and test the
cover delay, set the COVER switch back to OPEN po-
sition as shown in Figure 21.
ENTER
CASPER
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 21: Saving Cover Delay
•
To test, dip Casper into the application material until
the switching point is reached.
11.2
For Two Point / Pump Control Logic
•
Unscrew the cover and ensure that all DIP switches
are in OPEN position as shown in Figure 18. Make
sure that STATUS LED is not blinking for Error.
•
For setting cover delay for RELAY 1, follow the instruc-
tion given in section Single Point Switching.
•
For setting cover delay for RELAY 2, set the COVER
and HI CALIBRATE switch to CLOSE (Opposite of
OPEN for DIP switch) position as shown in Figure 22.
•
Press and hold ENTER key as shown in Figure 23.
The STATUS LED for RELAY 1 starts blinking. Blink
the LED according to value of cover delay.
•
Delay is entered, but not saved. To save and test the
cover delay, set the COVER and HI CALIBRATE switch
back to OPEN position as shown in Figure 24.
•
To test, dip Casper into the application material until
the switching point is reached.
ENTER
CASPER
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 22: Cover Delay Switch Position for Relay 2
ENTER
CASPER
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 23: Setting Cover Delay for Relay 2
ENTER
CASPER
ON
3
4
5
6
1
2
COVER
UNCOVER
CALIBRATE
HI CALIBRATE
FS1
FS2
RELAY 1
RELAY 2
SENS
ITIVITY
1
2
3
4
5
6
CLOSE
Lo
Figure 24: Saving Cover Delay for Relay 2
12
Uncover Delay
When the application material uncovers Casper’s probe,
the changeover of the output can be delayed by a pre-
determined time. This time is called UNCOVER Delay. For
a different value of uncover delay, the number of blinks can
be adjusted as per requirement.
The STATUS LED will start blinking RED if the switch point
is achieved. It will blink for the number of seconds for which
the uncover delay is set.
Note:
Set the value of UNCOVER DELAY between
1-25 secs.
12.1
For Single Point Switching
•
Unscrew the cover and ensure that all DIP switches
are in OPEN position as shown in Figure 18. Make
sure that STATUS LED is not blinking for error.
•
Set the UNCOVER switch to CLOSE (Opposite of
OPEN for DIP switch) position as shown in Figure 25.
•
Press and hold ENTER key as shown in Figure 26.
The STATUS LED for RELAY 1 starts blinking. Blink
the LED according to value of uncover delay.
Sapcon Instruments Pvt.Ltd.
®