28
SPECIFICATIONS
Name
Biomedical Freezer
Model
MDF-U730M
External dimensions
W770 x D830 x H1955 (mm)
Internal dimensions
W650 x D710 x H1520 (mm)
Effective capacity
690 L
Exterior
Painted steel
Interior
Painted steel
Door Painted
steel
Lock 1
Insulation
Rigid polyurethane foamed-in place
Shelf
Polyethylene coated wire
4 pieces
Size: W626 x D620 (mm) Max. load: 50 kg/shelf
Access port
Diameter 30 mm, 1 port on left side
1 port on the top of unit
Compressor
Hermetic type, 400 W
Evaporator
Tube on sheet type
Condenser
Wire and tube type
Refrigerant
R-404A
Temperature controller
Electronics controller
Temperature display
Digital display (between -41 and +1
o
C)
Temperature sensor
Thermistor sensor
Alarm & safety
High temp. alarm, Low temp. alarm, Power failure alarm, Door alarm,
Compressor temperature alarm,
Key lock, Thermal sensor abnormality
Memory backup
Nonvolatile memory
Weight 150
kg
Accessories
1 set of key,
1 scraper,
2 large & 2 small nylon clips (for temperature recorder)
Caster 4
Leveling foot
2
Optional component
Automatic temperature recorder (MTR-G85)+ Mounting kit (MPR-S7)
, Automatic temperature recorder (MTR-Mounting kit (MPR-S30)
Note
:
Design or specifications will be subject to change without notice.
The battery for power failure alarm is an article for consumption. It is recommended that the battery will
be replaced about every 3 years.