1
Comparison MDF-U53V with MDF-U52V
Item MDF-U52V
MDF-U53V
Remarks
Compressor
Sanyo
Toshiba career
Toshiba compressor has
been cooperated
development.
H stage side
1100W
1100W
L stage side
1100W
1100W
Filter alarm
FILTER lamp is lit
FILTER lamp is lit and buzzer
sounds intermittently
2.8yrs, battery accumulating
time
䋨
BATTERY lamp is lit
䋩
2.8yrs, battery accumulating
time
䋨
BATTERY lamp is lit
䋩
Input code ‘409’ in F06 to
turn BATTERY lamp off.
You can see accumulation
time in F03.
Notification of
parts change
Fan motor accumulation time,
none
5.6yrs, fan motor accumulating
time
䋨
BATTERY lamp is blinking
䋩
Input code ‘419’ in F06 to
stop BATTERY lamp
blinking.
You can see accumulation
time in F32.
Overload relay
Overload relay
Compressor
protection
䊶
H stage side compressor is
ceased when it detects 70C
higher than filter sensor
temperature.
䊶
E10 and PV are displayed
alternately.
䋨
Buzzer sounds and remote
alarm terminal turns over
䋩
䊶
H stage side compressor is
ceased when it detects 60C
higher than filter sensor
temperature.
䊶
E10 and PV are displayed
alternately.
䋨
Buzzer sounds and remote
alarm terminal turns over
䋩
When it detects 10C lower
than AT sensor
temperature, H stage side
compressor turns on.
Status 3
Adding test value, SV and AT
makes the compared value.
Compared value = 95
䋦䋨
fixed
䋩