19
F16:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F16” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F17:
<Purpose> Non-volatile memory initialization, model code change
<Operation> Service code should be input in F06 prior to use this mode.
Input F17 and press SET key to display “001”. Change the value
with key and key. Press SET key to store and revert to PV
display.
Model code ‘003’: MDF-C8V
F18:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F18” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F19:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F19” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F20:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F20” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F21:
<Purpose> Serial communication ID setting
<Operation> Input F21 and press SET key to display “000” (initial).
Change the value in the range “001”~”255” with key and key.
F22:
<Purpose> Serial communication mode setting
<Operation> input F22 and press SET key to display “000” (initial value)
Change the value with key and key.
Press SET key to store the value and revert to PV display.
Control mode (the 3
rd
digit)
0: Local (initial)
1: Remote
Baud rate (the 2
nd
digit)
0: 2400bps (initial)
1: 4800bps
2: 9600bps
Note) You cannot be changed SV if control mode is set in “Remote”.
F23:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F23” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F24:
<Purpose> To control remote alarm output
<Operation> Input F24 and Press SET key to display “000” (initial).
Change to the desired value with key and key.
Press SET key to store the value and revert to PV display.
000: Remote alarm is ON even if the buzzer stops beeping with BZ
key pressed.
001: Remote alarm is OFF when the buzzer stops beeping with BZ
key pressed.
Summary of Contents for MDF-C8V
Page 8: ...5 Dimensions 405 490 R 2 5 550 58 852 945 425 887 35 727 685 33 㧨 8㧪 ...
Page 9: ...6 Refrigeration circuit diagram ...
Page 12: ...9 Components on PCB ...
Page 15: ...12 Wiring diagram ...
Page 16: ...13 Circuit diagram ...
Page 36: ...33 MDF C8V Ultra Low Temperature Freezer INSTRUCTION MANUAL ...
Page 44: ...41 8 FREEZER COMPONENTS 13 12 5 11 1 2 3 6 7 8 5 4 10 9 ...
Page 76: ...73 CVK UB4 CVK UB4 I Back up System for Ultra Low Temperature Freezer INSTRUCTION MANUAL ...
Page 92: ...1 JK1659 150B ...