100
44
PERFORMANCE
Temperature control range
Ambient tempe5
o
C to 50
o
C (ambient temperature; 5
o
C to 35
o
C)
Temperature distribution
± 0.25
o
C* (ambient temperature; 25
o
C, setting; 37
o
C, 5%, no load)
Temperature variation
± 0.1
o
C (ambient temperature; 25
o
C, setting; 37
o
C, 5%, no load)
CO
2
control range
0 to 20%
CO
2
variation
± 0.15% (ambient temperature; 25
o
C, setting; 37
o
C, 5%, no load)
O
2
control range
1 to 18%, 22 to 80%
O
2
variation
± 0.2% (ambient temperature; 25
o
C, setting; 37
o
C, 5%, no load)
Chamber humidity
95 ± 5% R.H.
Maximum heat emission
740 kJ/h
Noise level
24 db (A scale)
Usable environment condition
Temperature; 5
o
C to 35
o
C, Humidity; equal or less than 80% R.H.
(The designed performance may not be obtained when
the ambient temperature is equal or less than 15
o
C)
Note:
The unit with CE mark complies with EC directives 89/336/EEC, 93/68/EEC and 73/23/EEC.