Sanxin EC5 Operation Manual Download Page 3

 

3. Conductivity Calibration 

3.1. Press          key to enter in conductivity measurement mode, rinse the probe in pure water 

and dry it.   

                                                                                 
3.2. Fill certain amount (about half volume of the calibration bottle) of 1413μS and 12.88mS 

standard solution in separate calibration bottles;   

 
3.3. Long press          to enter calibration mode; Short press          to exit. 

 
3.4. Dip the probe in 1413μS standard solution and stir gently, leave to stabilize 

(LCD displays        as shown in the picture on the right). Short press          to 

complete 1-point calibration. The tester then returns to measurement mode 

and calibration icon        will appears one the LCD.     

   
3.5. For multi-point calibration (either 2 or 3 point), rinse the probe in clean water and repeat step 

3.4  using  84uS  standard  solution  and  /  or  12.88mS  standard  solution.  After  completion  of  each 

point, the calibration icons        (for 84uS) and        (for 12.88mS) will appear on the LCD near the 

        . 

 

4. Conductivity Measurement 

4.1. Short press          to turn on the tester. Rinse the probe in pure water and dry it. Dip the probe 

in sample solution, stir gently, leave it to Stabilize, Get reading after LCD displays        .   

     
4.2. Press          to switch between parameter: 

COND

TDS

SAL

 

.   

 

5. Special Notes 

5.1. The tester uses 84μS, 1413μS and 12.88mS conductivity standard solution. User can conduct 2 

or 3 point calibration. Refer to the chart as follows. Generally, calibrating the tester with 1413μS 

conductivity standard solution alone will meet the testing requirement.   

Calibration Indication Icon 

Calibration Standards 

Measuring Range 

 

84μS/cm 

0 ~ 200μS/cm 

 

1413μS/cm 

200 ~ 2000μS/cm 

 

12.88mS/cm 

2 ~ 20mS/cm 

Diagram - 2 

 

Summary of Contents for EC5

Page 1: ...EC5 Conductivity Tester Conductivity TDS Salinity Temp Operation Manual Shanghai San Xin Instrumentation Inc ...

Page 2: ...ND TDS SAL 2 In mode setting short press to change parameter Unidirectional 1 Long press to enter calibration mode 2 In calibration mode short press to confirm calibration 3 When measured value is locked short press to unlock Pull the battery cap up Push the battery cover along to the direction of arrow Open the battery cover Insert the batteries 4 AAA and ensure the batteries polarity is as per d...

Page 3: ...ater and repeat step 3 4 using 84uS standard solution and or 12 88mS standard solution After completion of each point the calibration icons for 84uS and for 12 88mS will appear on the LCD near the 4 Conductivity Measurement 4 1 Short press to turn on the tester Rinse the probe in pure water and dry it Dip the probe in sample solution stir gently leave it to Stabilize Get reading after LCD displays...

Page 4: ...ute ammonia solution 1 88 C 5 4 TDS and conductivity is linear related and its conversion factor is 0 40 1 00 Adjust the factor in parameter setting P4 The factory default setting is 0 71 Salinity and conductivity is linear related and its conversion factor is 0 5 The tester only needs to be calibrated in conductivity mode then after calibration the meter can switch from conductivity to TDS or sal...

Page 5: ...con will display on LCD Press key to cancel auto automatic hold b Backlight P2 Off turn off backlight On turn on backlight 1 backlight will last for 1 minute c Temperature compensation factor P3 Adjust temperature compensation factor between 0 00 4 00 Factory default setting is 2 00 Details refer to clause 5 3 d TDS factor P4 Adjust TDS factor between 0 40 1 00 Factory default setting is 0 71 Symb...

Page 6: ...ed when instrument does not work well in calibration or measurement Calibrate and measure again after recover the instrument to factory default status 7 Technical Specifications Cond Measuring range 0 200 0μS 0 2000μS 0 20 00mS Resolution 0 1 1μS 0 01mS Accuracy 1 F S Calibration points 1 3points TDS Measuring range 0 100 0ppm 0 1000ppm 0 10 00ppt Resolution 0 1 1ppm 0 01ppt Salinity Measuring ran...

Page 7: ...ff the screw ring unplug the probe plug in new probe pay attention to the probe s position and screw on the screw ring 10 Warranty We warrant this instrument to be free of defects in parts and workmanship for two years from date of shipment a six month limited warranty applies to probes This warranty does not apply to defects resulting from actions of the user such as misuse improper wiring operat...

Page 8: ...Shanghai San Xin Instrumentation Inc Add 4 F Building No 16 No 481 Guiping Road Shanghai China 200233 Tel 86 21 63362480 Fax 86 21 64956880 Website www shsan xin com E mail wxmab shsan xin com ...

Reviews: