DRAWING No. RD700/701 06-1404 2040 6010
V/ADP/
Ω
/ / /
/TEMP
/Hz
•
COM
DC•AC 400 mA
— 2 —
20 VAC rms
Hz
µA•mA
Ω
• •
• TEMP
0.63 A/500 VFuse
IR 50 kA
DC•AC 10 A
(10 A continuous)
A
V
ADP
Functions
Input
terminals
Maximum rating
input value
Maximum overload
protection input
1-3 Overload protections
DC•AC 1000 V
DC•AC 400 mV
1050 Vrms, 1450 Vpeak
600 VDC/AC rms
Voltage and
Current input
prohibited
µA / mA
•
COM
A
•
COM
12.5 A/500 VFuse
IR 20 kA
1
5. Never use meter for measuring the line connected with
equipment (i.e. motors) that generates induced or surge
voltage since it may exceed the maximum allowable voltage.
1
6. Never use uncased meter.
1
7. Be sure to use a fuse of the specified rating or type. Never use a
substitute of the fuse or never make a short circuit of the fuse.
1
8. When connect and disconnecting the test leads, connecting
the ground lead (black one) first. When disconnecting them,
the ground lead must be disconnected last.
1
9. Always keep your fingers below the finger guards on the
probe when making measurements.
10. Be sure to disconnect the test pins from the circuit when
changing the function.
11. Before starting measurement, make sure that the function and
range are properly set in accordance with the measurement.
12. Never use meter with wet hands or in a damp environment.
13. Do not use the device near an item of strong electromagnetic
generation or a charged item.
14. Never open tester case except when replacing batteries or
fuse. Do not attempt any alteration of original specifications.
15. To ensure safety and maintain accuracy, calibrate and check
the tester at least once a year.
16. The multimeter is restricted to indoor use only.