![Sanwa CD771 Instruction Manual Download Page 46](http://html.mh-extra.com/html/sanwa/cd771/cd771_instruction-manual_1198190046.webp)
— 15 —
Function
Input Terminal
Built-in Fuse
0.5A/1000V Fuse
Breaking capacity 30kA
Po
wer supply
Load
µA · mA measurement
and
Po
wer supply
Load
Function
Input Terminal
Built-in Fuse
10A/1000V Fuse
Breaking capacity 30kA
A measurement
10A
and
• If the indication will change little when an input signal is applied or a
current value which is significantly smaller than the expected value
is indicated, possible causes are the input terminals, incorrect
setting of the function switch, or blown fuse. Check these places.
Remarks:
• Maximum 20A can be measured if
the measurement time is less than
10 seconds.
(Take 10minutes or longer intervals
between measuremtents.)
7
DRAWING No. CD771 03-1101 2040 6017
CD771̲E 11.1.11 18:28 ページ15
Summary of Contents for CD771