11. What if there are error messages?
Error
message
Cause
Solution
ER1
The ear thermometer
is not yet ready for
measurement.
Take measurements only
once body ( ) or object
temperature mode (
)
appears on the display.
ER2
The measured ambi-
ent temperature is not
in the range from 10 °C
– 40 °C (50 – 104 °F).
Place the ear thermometer
in a room with an ambient
temperature between 10 °C
and 40 °C (50 °F and 104 °F)
for at least 30 minutes.
ER3
The ear thermometer
is not placed in the ear
correctly.
Please observe the notes
in the chapter “Measuring”.
Perform the measurement
again.
ER4
The ear thermometer
registered a large
change in the ambient
temperature.
Place the ear thermometer
in a room with an ambient
temperature between 10 °C
and 40 °C (50 °F and 104 °F)
for at least 30 minutes.
Error
message
Cause
Solution
ER5
The ear thermom-
eter is not working
properly.
Replace the batteries and
try again. If the fault occurs
again, contact Customer
Services.
ER6
You are in the “Measur-
ing object temperature”
mode and the protec-
tive cap is still on.
The protective cap must be
removed to measure object
temperatures.
In-ear measurement:
the measured tem-
perature is over 43 °C
(109.4 °F).
Object measurement:
the measured temper-
ature is over 100 °C
(212 °F).
Ensure that the device and
the protective cap are fully
functioning. Perform the
measurement again.
Lo
In-ear measurement:
the measured temper-
ature is below 34°C
(93.2°F).
Object measurement:
the measured
temperature is below
-20 °C (-4 °F).
Ensure that the device and
the protective cap are fully
functioning. Perform the
measurement again.
The batteries are
empty.
Remove the two empty
batteries and insert two
new 1.5 V AAA batteries in-
to the ear thermometer.
Summary of Contents for SFT 53
Page 59: ...59 Sanitas...
Page 60: ...60 4 1 2 3 4 5 1 2 3 AM PM 4 5 6 7 C F 8 9 10 11 12 5 1 2 3 4 5 E 1 5 7 6 8 9 10 11 12 2 3 4...
Page 61: ...61 6 6 1 1 24 H 2 12 24 3 4 5 6 7...
Page 62: ...62 6 2 C F C F 1 C 2 C F 7 30 7 1 6 6...
Page 63: ...63 20 1 2 3 4 5 38 C 100 4 F 37 5 37 9 C 38 C 100 4 F...
Page 64: ...64 7 2 1 2 3 4 1 2 5 7 3 3...
Page 65: ...65 7 4 10 1 2 3 M 3 4 3 8 1 2...
Page 66: ...66 3 1 5 AAA 9 10 Pb Cd Hg WEEE Waste Electrical and Electronic Equipment...
Page 69: ...69 13...