TROUBLESHOOTING

 Unplug unit before servicing 

 

Symptom 

Possible Cause 

Remedy 

Unit does not turn on 

Power cord unplugged 
No power at outlet 

Plug HFD cord in outlet. 
Check voltage at outlet (115VAC) 

HFD circuit breaker off 

Turn on breaker.  HFD6 & 7: remove 
lower rear cover to access breaker. 

Internal power cord 
disconnected 

Check resistance across plug, 
ensure the breaker is on, measure 
resistance across plug terminal, line 
and neutral, should be about 160

Ω.

 

Front glass door open 

HFD6 & 7 only:  Controller should 
rea

d “OPEn” . Shut front glass door, 

press power button on controller to 
initialize. 

One or more Xenon lamps 
is off 

Bulb blown 

Check at terminal for resistance of 
about 44

Ω

Check to ensure bulb is securely 
installed. 
Replace blown bulb(s). 

HFD display ambient does 
not reach 150°F 

One or more heater 
plates not reaching 
temperature. An error 
(E70 to E75) may be 
displayed on controller. 

HFD, HFDC and HFDM need 30 
minutes to reach temperature.  
HFDH and HFDHC needs at least 1 
hour.  
Check plate temperature at controller 
(see controller operation). 

One or more Xenon 
lamp not functioning 

Check at terminal for resistance of 
about 44

Ω

Check to ensure bulb is securely 
installed. 
Replace blown bulb(s). 

E60 to E65 error 

Heater thermistor open 
circuit 

Pull out controller, disconnect 
corresponding heater connector.  If 
resistance between the red wires is 
>100k

Ω

, check cause of open circuit.  

Heater may need replacing. 

Heater plate not heating 

Fuse blown on controller  Remove controller and check fuses. 

Wires disconnected 

Remove controller and check all 
connections at controller. 

Failed heating element 

Check resistance across white and 
black terminals from heater.  
Resistance should be about 95

Ω

 at 

room temperature. 

 

 

 

Summary of Contents for Vendo HFD000006

Page 1: ...Service Manual MODELS HFD000006 7 HFDC00002 HFDM00002 HFDH00001 HFDHC0001 2 3 4 Table of Contents Commitment to Safety 2 Safety Warnings 3 Parts 4 Operation 8 Troubleshooting 12 Controller Connection...

Page 2: ...effective safety program Use these precautions along with the service manual when installing or servicing the HFD We strongly recommend a similar commitment to safety by every servicing organization O...

Page 3: ...lectric shock or fire do not use volatile or flammable products in or near the HFD When moving the HFD turn the HFD off unplug and allow the unit to cool To prevent glass breakage and injury do not pu...

Page 4: ...6 Shelf Assembly Middle Right 10 1207453 Shelf Assembly Upper Right 11 1158685 Gasket A 12 1158696 Gasket B 13 1159513 Lamp Socket 14 1159525 Xenon Heating Lamp 15 1216614 Power Cord Plug 16 1161921 C...

Page 5: ...Vertical 2x 5 1187405 Damper one each side 6 1202273 Extension Spring one each side 7 1199716 Rack Top Middle Shelf not shown 4x 8 1199728 Rack Bottom Shelf not shown 2x 9 1201311 Shelf Guard 4x 10 12...

Page 6: ...98 Front Door 1203125 Caution Hot Label 1182055 Middle Upper Shelf Rack Black 1158685 Side Gasket 1178052 Horizontal Rear Gasket 1166955 Side Glass no decal 1165366 Right Decal 1201293 Controller 1178...

Page 7: ...asket 1242342 Pan Restraint 2x See last page for Controller Connections for electrical component part numbers 1185603 4 Vertical Frame Gasket 1185603 3 Horizontal Frame Gasket 1226156 Left Front Door...

Page 8: ...30 minutes again Once the timer is activated the LED next to the shelf timer button becomes green The display then reverts to HFD internal ambient temperature display While a timer is on pressing a t...

Page 9: ...back on the lights in the HFD will come on Error Codes E30 Internal ambient thermistor open circuit E60 Upper left shelf thermistor open circuit except HFDM E63 Upper right shelf thermistor open circ...

Page 10: ...0 UR 235 1 00 1 00 MR 235 2 00 2 00 LR 220 2 00 1 00 CF 1 DA 1 DS 1 SV01 001n UL 000 2 00 2 00 ML 000 2 00 2 00 LL 245 2 00 2 00 UR 245 2 00 2 00 MR 245 2 00 2 00 LR 245 2 00 2 00 CF 1 DA 1 DS 0 SV01...

Page 11: ...2 00 2 00 LL 245 2 00 2 00 UR 245 2 00 2 00 MR 245 2 00 2 00 LR 245 2 00 2 00 CF 1 DA 1 DS 0 HFDH and HFDHC Left Side Right Side Heater Setpoint F Left Timer Right Timer Heater Setpoint F Left Timer R...

Page 12: ...lown bulb s HFD display ambient does not reach 150 F One or more heater plates not reaching temperature An error E70 to E75 may be displayed on controller HFD HFDC and HFDM need 30 minutes to reach te...

Page 13: ...HC 1202224 Remote on off cable 1158908 Thermistor 1206552 Door Switch Cable 1206564 Lamp Cable 6x HFD HFDC 1206564 Lamp Cable 3x HFDM 1232444 Lamp Cable 6x HFDH HFDHC 1159513 Lamp Socket 6x HFDM 3x 11...

Reviews: