9
BP 3ABO-H
EN
* Please consult your doctor, if this or any other problem occurs
repeatedly.
8. Safety, Care, Accuracy Test and Disposal
Safety and protection
«
ERR 2
» Error signal
During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be generated
in the cuff. A leak may have occurred. Check
that the cuff is correctly connected and is
not too loose. Replace the batteries if
necessary. Repeat the measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and
no result can therefore be displayed. Read
through the checklist for performing
reliable measurements and then repeat
the measurement.*
«
HI
Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
«
LO
»
Pulse too low The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*
Error
Description Potential cause and remedy
If you think the results are unusual, please read through
the information in «Section 1.» carefully.
•
This instrument may be used only for the purpose described
in this booklet. The manufacturer cannot be held liable for
damage caused by incorrect application.
•
This instrument comprises sensitive components and must be
treated with caution. Observe the storage and operating
conditions described in the «Technical Specifications» section!
•
Protect it from:
−
water and moisture
−
extreme temperatures
−
impact and dropping
−
contamination and dust
−
direct sunlight
−
heat and cold
•
The cuffs are sensitive and must be handled with care.
•
Only pump up the cuff once fitted.
»