35
SAMIX GmbH
Fault causes/troubleshooting
8
Fault causes/troubleshooting
8.1
Instructions in the event of faults
Fault
Troubleshooting
The SAMIX
®
BR200 cannot be
switched on.
Check whether there is supply voltage and the connector is correctly
connected to the socket.
The SAMIX
®
BR200 switched off
due to an overload.
Set the adjusting knob to "0".
Observe a cooling period of 30 minutes.
Tab. 3: Faults and troubleshooting
After the device is switched off and the fault has been eliminated, the SAMIX
®
BR200
works again normally.
Following troubleshooting perform a function check in accordance with the details
under "Function check or initial operation" in the enclosed document "Installation
Qualification (IQ)".
In the case of serious faults, note the device type and contact the manufacturer.
NOTICE
In the case of serious faults, only send in the drive head of the SAMIX
®
BR200 to SAMIX
GmbH. Unscrew the drive head by slackening the wing bolt on the bottom of the device foot
NOTICE
After an operation period of approximately 500 hours, the contact carbon brushes of the
SAMIX
®
BR200 may be worn. The contact carbon brushes are replaced during the regular
maintenance work performed by SAMIX GmbH.
The inspection and maintenance work to be performed for documentation within the frame-
work of a QM system can be obtained from customer service upon request.