4.2 Clean the optical fiber under test and connect to the OTDR interface of the instrument. Pay

attention to the type of the optical fiber street under test. If the mismatched connector type is

connected to the instrument, the output optical fiber head of the instrument may be damaged and the

correct measurement result may not be obtained.

4.3 Press the SCAN key, the instrument begins to SCAN the optical fiber, and the scanning progress

is displayed on the screen. The instrument does not need to set pulse width, and is automatically

adjusted in the scanning process. Based on powerful intelligent analysis software, the optical fiber

event points in the whole measurement range are calculated.

4.4 The measurement is completed and all measurements are displayed. The total number of events

can be up to 30, press the up and down key to scroll through the value of each event. The error of

reflection event is small, and the error of attenuation event is relatively large.

4.5 An empty event is displayed if the measured value is in or out of the instrument blind area.

4.6 The position of the current breakpoint is displayed after instrumentation is completed.

Attention: Do not look directly at the end or break of the fiber or tail under test to prevent eye

damage.

5. Standard configuration

Hand-held portable Optical Fiber Ranger, instructions for use, USB charging head, USB charging

cable, Oxford soft pack, SC adapter, packaging carton.

Note: lithium battery charging voltage is 5V.Mobile phone charger can be used.

6. Common Troubleshooting Solutions:

Failure Phenomenon

Possible Cause

Terms Of Settlement

The instrument can not

start

The power is not open

Press switch key

Insufficient electricity

Charging in time

Incorrect measurement

of optical fiber length

Incorrect refractive index

setting

Set the refractive index of light to the

refractive index calibrated by the

manufacturer of light

Pollution of Optical Fiber Ends

Wipe the end face of optical fiber with

absolute ethanol

Contamination of optical fiber

end face in optical output

connector of instrument

Clean with absolute ethanol