Tel: 886.2.2175 2930 Email: [email protected]
54
6.3
Example 3: Analyzing Input Signals to Eliminate Random Noise
In certain circumstances, to display a noisy signal on the oscilloscope and to get its details, you may follow the steps
below to analyze this signal.
Observing Noisy Signal
1.
Press the ACQUIRE button to see the Acquire menu.
2.
Push the Type option button and select Real Time.
3.
Push the Peak Detect option button.
4.
If necessary, push the DISPLAY button and set the Contrast option to view the noise more clearly.
See the figure below for better understanding.
Eliminating Random Noise
1.
Press the ACQUIRE button to see the Acquire menu.
2.
Push the Type option button and select Real Time.