Document Reference: ENG-UMAN-000730
Revision: 6.00
Page 12 of 40
5
Percutaneous Lead Implant Procedure
Overview
Figure 5.1: Overview of the percutaneous lead implantation process
Percutaneous Lead Placement
Insert the epidural needle
1.
Mark the patient’s back using fluoroscopy for the required vertebral level.
Warning:
The risk of patient injury increases as the needle insertion site
moves up the spinal column from lumbar to cervical.