11. TROUBLESHOOTING (2)
PROBLEM
POSSIBLE CAUSE
SOLUTION
LCD shows low battery symbol
LCD shows low battery symbol
Change the batteries
LCD shows “Er 0”
Pressure system is unstable before
measurement
Don’t move and try again.
LCD shows “Er 1”
Fail to detect systolic pressure
LCD shows “Er 1”
Fail to detect diastolic pressure
LCD shows “Er 1”
Pneumatic system blocked or cuff is
too tight during inflation
Apply the cuff correctly and try again
LCD shows “Er 4”
Pneumatic system leakage or cuff is
too loose during inflation
LCD shows “Er 5”
Cuff pressure above 300mmHg
Measure again after five minutes. If
the monitor is still abnormal, please
contact the local distributor or the
factory.
LCD shows “Er 6”
More than 3 minutes with cuff
pressure above 15 mmHg
LCD shows “Er 7”
EEPROM accessing error
LCD shows “Er 8”
Device parameter checking error
LCD shows “Er A”
Pressure sensor parameter error
No response when you press button
or load battery.
Incorrect operation or strong
electromagnetic interference
Take out batteries for five minutes,
and then reinstall all batteries.
NOTICE
0
1. Do not drop this monitor or subject it to strong impact.
0
2. Avoid high temperature and solarization. Do not immerse the monitor in water as this will result in damage to the
monitor.
0
3. If this monitor is stored near freezing, allow it to acclimate to room temperature before use.
0
4. Do not attempt to disassemble this monitor.
0
5. If you do not use the monitor for a long time, please remove the batteries.
0
6. It is recommended the performance should be checked every 2 years or after repair. Please contact the service center.
0
7. Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after moistened with water, diluted disinfectant
alcohol, or diluted detergent.
0
8. No component can be maintained by user in the monitor. The circuit diagrams, component part lists, descriptions,
calibration instructions, or other information which will assist the user’s appropriately qualified technical personnel to
repair those parts of equipment which are designated repairably can be supplied.
0
9. The monitor can maintain the safety and performance characteristics for a minimum of 10,000 measurements or three
years, and the cuff integrity is maintained after 1,000 open–close cycles of the closure.
10. It is recommended the cuff should be disinfected 2 times every week if needed (For example, in hospital or in clinique).
Wipe the inner side (the side contacts skin) of the cuff by a soft cloth squeezed after moistened with Ethyl alcohol
(75-90%), then dry the cuff by airing.