PRODUCT DESCRIPTION
Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse rate can be
measured automatically and non-invasively. The LCD display will show blood pressure and pulse rate. The most recent
2x60 measurements can be stored in the memory with date and time stamp. The monitor can also show the average
reading of the last three measurements.The Electronic Sphygmomanometers corresponds to the below standards: IEC
60601-1:2005/EN 60601-1:2006/AC:2010 (Medical electrical equipment -- Part 1: General requirements for basic safety
and essential performance), IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment -- Part 1-2:
General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -
Requirements and tests), IEC 80601-2-30: 2009+Cor.2010/EN 80601-2-30:2010(Medical electrical equipment –Part 2-30:
Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers),
EN 1060-1: 1995 + A1: 2002 + A2: 2009 (Non-invasive sphygmomanometers - Part 1: General requirements), EN 1060-3:
1997 + A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-
mechanical blood pressure measuring systems).
SPECIFICATIONS
Product name:
Blood Pressure Monitor
Model
BPA-9200-GB
Classification
Internally powered, Type BF applied part, IPX0, No AP or APG,
Continuous operation
Machine size
Approx. 138mm × 98mm × 48mm (5 7/16”x 3 27/32”x 1 7/8”)
Cuff circumference
30cm-42cm (11 13/16”-16 17/32”), 22cm-30cm (8 21/32”-11
13/16”) (Optional), 42cm-48cm (16 17/32”-18 29/32”)
(Optional)
Weight
Approx. Approx. 211g (7 7/16 oz.) (exclude batteries and cuff)
Measuring method
Oscillometric method, automatic inflation and measurement
Memory volume
2x60 times with time and date stamp
Power source
Batteries: 4 ×1.5V
Size AA
Measurement range
Cuff pressure: 0-300mmHg
Systolic: 60-260mmHg
Diastolic: 40-199mmHg
Pulse rate: 40-180 beats/minute
Accuracy
Pressure: ±3mmHg
Pulse rate: ±5%
Environmental temperature for operation
10ºC~40ºC (50ºF~104ºF)
Environmental humidity for operation
≤90%RH
Environmental temperature for storage and transport
-20ºC~55ºC (-4ºC~131ºC)
Environmental humidity for storage and transport
≤90%RH
Environmental pressure
80kPa-105kPa
Battery life
Approx 500 times
All components belonging to the pressure measuring
system, including accessories
Pump, Valve, LCD, Cuff, Sensor
Note:
These specifications are subject to change without notice.