8 |
GB
GB
| 9
CLINICAL ACCURACY VALIDATION METHOD
This device is an adjusted mode clinical thermometer
the validated information for clinical accuracy in each adjusted mode are:
GroupA1:
∆cb=-0.01°C, L
A
=0.18,
δ
r = ±0.08°C
GroupA2:
∆cb= 0.06°C, L
A
=0.22,
δ
r = ±0.08°C
GroupB :
∆cb=-0.01°C, L
A
=0.20,
δ
r = ±0.07°C
GroupC :
∆cb=-0.01°C, L
A
=0.18,
δ
r = ±0.07°C
∆cb: CLINICAL BIAS
L
A
: LlMITS OF AGREEMENT
δ
r: CLINICAL REPEATABILITY
ERROR CODES
When a malfunction or incorrect temperature measurement occurs, an error message will
appear as described below.
LCD Display
Cause
Solution
The temperature measured
is lower than 34°C (93.2°F)
Operate the thermometer only
between the specified temperature
ranges. If necessary, clean the
sensor tip. In the event of a
repeated error message, contact
your retailer or Customer Services.
The temperature measured
is higher than 43°C
(109.4°F)
The operating temperature
is not in the range
15°C~35°C (59°F~95°F)
Operate the thermometer only
between the specified temperature
ranges.