Model WS / SBI-140 Indicator
User Manual
18
Service Settings Continued
Parameter
Settings
Default settings in bold
P12.x
Power up Zero range
Selects the Power up zero-point range based on the
calibration zero point.
P12.0 = +1%
P12.1 = +2%
P12.2 = +5%
P12.3 = +10%
P12.4 = +20%
P12.5 = +50%
P12.6 = +100%
P12.7 = No limitation
P13.x
Zero button range
Selects the zero range the zero buttons can zero off.
P13.0 = ±1% P13.8 = +1%
P13.1 = ±2% P13.9 = +2%
P13.2 = ±5% P13.10 = +1%
P13.3 = ±10%
P13.4 = ±20%
P13.5 = ±50%
P13.6 = ±100%
P13.7 = No limitation
P14.x
Scale Power up when inside the power up zero range
Allows the scale to power up and zero from the following
point.
0= Power up and zero at any weight.
1= Power up zero based off calibration zero-point.
2= Power up back at the zero-point the scale was
powered off at and also display any active tare.
P14.0
P14.1
P14.2
P15.x
Scale Power up when outside the power up zero range
Allows the scale to power up and display the following:
0= Zero and display in current Gross weight.
1= Displays Gross weigh
t
based off calibration zero point.
2= Displays Gross or Net weight based off the zero-point
when the scale was last powered Off.
3= Continuously display error message “0¯¯¯¯”
P15.0
P15.1
P15.2
P15.3
P16.x
Zero tracking range
P16.0 = off P16.5 = ±2d
P16.1 = ±0.25d P16.6 = ±3d
P16.2 = ±0.5d P16.7 = ±4d;
P16.3 = ±1d
P16.8 = ±5d
P16.4 = ±1.5d
P17.x
Data filter intensity
P17.0 = very weak
P17.1 = weak
P17.2 = middle
P17.3 = strong
P18.x
Check weight stability range
P18.0 = ± 0.5d; P18.5 = ± 4d
P18.1 = ± 1d
P18.6 = ± 5d
P18.2 = ± 1.5d P18.7 = ± 6d
P18.3 = ± 2d P18.8 = ± 7d
P18.4 = ± 3d P18.9 = ± 8d
P19.x
Overload limit range
P19.0 = 0 P19.5 = 110%
P19.1 = + 9d
P19.6 = 120%
P19.2 = 101% P19.7 = 150%
P19.3 = 102% P19.8 = 200%
P19.4 = 105 % P19.9 = No limitation
Note: Default settings may vary depending on platform configuration required.