Salter Brecknell S-100 Operation Manual Download Page 10

User Configuration Settings 

Parameter 

 

Setting 

Default settings  in bold

P1.xy 

Auto shutoff  timer in minutes

  

Set up time for the auto off function. 

(00 = 0ff, 01-15 = time in minutes) 

P1.00  
P1.01 

 

– P1.15,  

P1.5 = 15 minutes 

P2.x 

Hold and print key functionality 
Setup button function 

0 = Press button once to activate hold 
1=  Press button once to print 
2=  Press button to print/Press and hold button to activate hold. 

P2.0= Hold  
P2.1= Print;  

P2.2= Print & Hold,  

P3.xy 

Hold Function Settings 

0= No hold function active. 
 
1= Averaging hold with manual push button release. 
The weight reading will be held on the display until a higher 
weight is applied, this will automatically release the held weight 
and re-hold it at the new higher weight reading.         
 
2= Averaging hold with automatic release and re-hold.  
As above but The weight reading will be held on the display until 
the platform is empted and the next weight reading over 
10divisions is applied.  
  
3-50= Selectable hold window from  +/- 3  to 50 divisions 
Will hold display reading once stable within a selectable weight     
range, to release the hold button must be represses.. 

P3.0 
P3.1 

P3.2 

P3.3 to 50 

P4.x 

RS232 – Serial Interface 
Settings for serial interface 

0= No RS232 output  
 
1= Print displayed data once stable when print key is pressed;  
 
2= Print gross, tare and net weight  once stable when print key is 
     pressed  
 
3= Continuously output gross weight  
 
4= Continuously output gross, tare and net weight  
      
 
5= Print displayed data once stable one time only.  
 
6= Print gross, tare and net weight once stable, one time only.   
 
7= Bidirectional - RS232, SBI protocol 

P4.0                    
P4.1                     

P4.2                    

 

P4.3                    
P4.4 
P4.5  
P4.6 
P4.7 

P5.x 

RS232 Baud rate  

P5.0= 1200

   P5.3= 9600

 

P5.1= 2400   P5.4= 19200
P5.2= 4800 

P6.x 

RS232 Data format  

0 = 8 digits,  no odd or even,    1 start bit, 1 stop bit 
1 = 7 digits, 1 even, 1 start bit , 1 stop bit 
2 = 7 digits, 1  odd,  1 start bit,  1 stop bit 
3 = 8 digits, no parity, 1 start bit, 1 stop bit 
 

P6. 0  
P6. 1 
P6. 2 

P6. 3 

P7-P19 .x 

SERVICE CONFIGURATIONS ONLY 

 

Any adjustment to these settings could seriously affect the indicators performance.  

Seek advice from a service engineer before changing. 

 

 

Summary of Contents for S-100

Page 1: ...Model S 100 Floor Scale Operation Manual Manual xxxxxx Issue xx...

Page 2: ...DOUBT SEEK ADVICE FROM A QUALIFIED ELECTRICIAN Routine maintenance To avoid the possibility of electric shock or damage to the machine always switch off the machine and isolate from the power supply...

Page 3: ...nu 14 Service Configuration Settings 15 Technical Specifications 16 Error Messages 16 Declarations of Compliance 17 Contact Information Technical Support 17 Product Identification Table Part Number Ca...

Page 4: ...oadcell socket Power Adapter Socket Power Supply Connecting the Indicator to the Platform 6VDC 500mA 230V 50Hz Wire the cable attached to the base as shown if necessary 4 2 3 1 1 RED 2 RED 3 RED 4 RED...

Page 5: ...Indicator and Platform Assembly Rem ove packagin g pa ds x2 x2 1 2 3 x4 4 5...

Page 6: ...0 Weight is zero On Off Zero On Off Zero kg lb Unit of measure Flashing LED means that the weight reading is not stable Units kg lb Unit of measurement NET Tare is activated Hold Print Hold Print CH B...

Page 7: ...tion Turning On and Zeroing the Scale On Off Zero ZERO 0 Turning off the scale On Off Zero Selecting Unit of Measure Units kg lb Using the Tare 1 2 3 NET 4 NET Removing the Tare On Off Zero 0 ZERO kg...

Page 8: ...old function with manual release This function needs to be setup in Parameter P3 1 shown on page 10 1 Hold Print Hold 2 3 Removing Hold Hold Print Hold Print Function For communications to a printer o...

Page 9: ...ecting parameter Tare Changing data within the parameter Units Saving data Tare Exiting setup On Off Zero ZERO 0 Gross 45 00kg Tare 1 35kg Net 43 65kg Hold Print kg Moves flashing digit Units Change f...

Page 10: ...ctable hold window from 3 to 50 divisions Will hold display reading once stable within a selectable weight range to release the hold button must be represses P3 0 P3 1 P3 2 P3 3 to 50 P4 x RS232 Seria...

Page 11: ...tare LF H1H2H3 CR ETX U CR Changes the units LF u1u2 CR LF H1H2H3 CR ETX L CR Activates the hold function LF H1H2H3 CR ETX X CR Switches off the scale Indicator switches off Unrecognised command LF CR...

Page 12: ...minus decimal point unit CR EXT LF Tare reading decimal point unit CR EXT LF Net reading minus decimal point unit CR EXT RS232 serial interface wiring DE 9 Female Scale DE 9 Male Host Pin Name Direct...

Page 13: ...bration weight 0 KG 1 LB Factory Calibration Settings shown in bold Primary Capacity Displayed Resolution Units Selectable Capacity P7 P8 P9 P10 150 lb x 0 02 lb 7500 75 kg x 0 01kg 16 1 2 1 150 x 0 0...

Page 14: ...g lb 5 Single point calibration enter the same weight in again and move to number 7 For 2 point calibration enter in the second calibration weight between 25 100 full capacity Tare 6 7 Tare 8 9 Tare Z...

Page 15: ...5 50 6 100 7 no limitation P13 00 15 3 Zero button range Full scale 00 1 06 20 12 5 01 2 07 no limit 13 10 02 3 08 1 14 20 03 4 09 2 25 no limit 04 5 10 3 05 10 11 4 P14 0 1 2 0 Signal within power on...

Page 16: ...ill be turned off in two minutes automatically AC Adapter 10 5 12VDC 600mA with central positive 5 7 VDC 9 VDC Error Messages Error Message Definition Required Solution Troubleshooting 0 Weight above...

Page 17: ...ications Le pr sent appareil num rique n met pas de bruits radio lectriques d passant les limites applicables aux appareils num riques de la Classe A prescrites dans le R glement sur le brouillage rad...

Reviews: