Pegasus Floor Scale Installation Guide

1. Make sure deck is empty and indicator is zeroed.

2. Load deck with as much evenly distributed test weight as available (not to

exceed scale capacity).

3. Unload deck and check for zero shift.

4. Rezero indicator if necessary and reload deck.

5. If necessary, trim the FINE SPAN control in the indicator for an indication

precisely equal to the calibrated test weights applied to the deck. See
your indicator's Service Manual for details on this procedure.

6. Enable AZT on indicator.

Your scale is now corner balanced and the system is calibrated.

Final Span Calibration

To replace one or more weight sensor follow this procedure:

1. Obtain access to junction box and the junction box cover.

2. Disconnect weight sensor cable from junction box terminal bar.

3. Set scale on blocks so feet are clear of the ground.

4. Unscrew foot from weight sensor(s) to be replaced.

5. Remove weight sensor mounting bolts and hardware.

6. Remove weight sensor.

7. Install new weight sensor.

8. Reinstall mounting hardware for weight sensor. Torque to 85 ft/lbs.

9. Replace foot.

10. Re-route weight sensor cable to junction box and connect.

11. Lower scale to floor and level.

12. Check corner balancing and calibration of the scale. Adjust if necessary.

13. Replace J-box cover.

Weight Sensor Replacement

Span calibration is not required
on precalibrated systems.

Summary of Contents for Pegasus Floor Scale

Page 1: ...reas and these areas should not give under loads Figure 1 Pegasus scales This scale is precalibrated and is ready for installation and use Place the Pegasus deck on a solid level surface Use the level...

Page 2: ...Yellow Sense White Signal Orange White Shield Red Signal Blue Sense Black Excitation Remove the access plate and the junction box cover plate Attach the leads of the indicator interface cable to TB5...

Page 3: ...he value of internal zero for your particular indicator refer to yourindicator sServiceManual 3 Use test weights equal to 20 25 of full capacity and obtain a displayed weight value for the test weight...

Page 4: ...ibrated Final Span Calibration To replace one or more weight sensor follow this procedure 1 Obtain access to junction box and the junction box cover 2 Disconnect weight sensor cable from junction box...

Page 5: ...y scales You can use shim stock under these floor plates to obtain a level mounting surface Drill the four corner holes through the shim stock but leave the center intact These drawings are to scale Y...

Page 6: ...6 Pegasus Floor Scale Installation Guide 10K Scales Floor Plate Part Numbers SST 28273 0012 1 4 thick SST 28273 0020 1 2 thick...

Page 7: ...774 0011 Junction Box ABS plastic 1 1 41244 0018 Foot Assy 4 41244 0026 Foot Assy 4 53455 0017 Foot Assy 4 53455 0033 Foot Assy 4 52655 1064 Bolts Hex Cap 20 x2 8 8 22408 1034 Bolts 20 x1 75 8 22408 1...

Page 8: ...eigh TronixInc 03 24 03 PEGASUS_i P65 PN43008 0010Ae1 PrintedinUSA P O Box 9533 3620 Central Ave NE Minneapolis MN 55418 Toll free 800 637 0529 Tel 612 781 4211 Fax 612 781 4320 e mail sales salterbre...

Reviews: