PC-150 Operation Manual

   

9

Setup Menu Mode / 

Continued

SETUP MENU DESCRIPTIONS

NAME/CODE

DESCRIPTION

CODE/VALUE

F1

Display Mode

Selects display mode for the scale.

"0" = Non-NTEP mode

"1" = NTEP mode

0

√√√√

1

F6

Zero Range

Selects the range within which the  scale  may  be  zeroed.  Note  that

the scale must be in standstill to zero the scale.

100%

√√√√

1.9%

F9

Calib. Unit

Selects the primary base unit to be used in the calibration process.

Also the default unit for normal operation.

"0" = primary unit is lb.                       "1" = primary unit is in kg.

0

√√√√

1

F11

Display Check

Actuates  the  function  that  illuminates  all  digit  segments,  decimal

points,  and  LCD  annunciators  in  a  test  sequence.  Pressing  the

ZERO

 key to scroll down one level begins the test sequence.

Press 

ZERO

 key

to begin sequence

F12

Calibration

Places  scale  into  the  calibration  routine.  Scrolling  down  with  the

ZERO

 key one level begins the procedure.

Press 

ZERO

 key

to begin sequence

F13

Factory Reset

This sub-menu will reset all parameters in the “F” and “A” menus to

the default settings. USE WITH CAUTION!!!!!

Press the 

ZERO

key twice to exe-

cute.

NOTE: 

S

ub-menu F13 restores all parameters in the User (“A”) Menu and Setup (“F”) Menu

to the factory default settings. PLEASE USE CAREFULLY AS YOU MAY LOOSE VITAL

SETUP PARAMETERS!!!

Calibration

ON/OFF

SWITCH

DC POWER

JACK

RS-232

PORT

ID TAG

CALIBRATION

SWITCH COVER

CALIBRATION MODE KEY FUNCTIONS

ZERO

GROSS

NET

TARE

PRINT

SET

lb

kg

The scale can be calibrated by following the procedure below. The minimum test weight that can

be used is 1.5 lb (0.7 kg), but 100 lb (50 kg) is the recommended test weight.

Summary of Contents for PC 150

Page 1: ...on 1 2 September 14 2000 Contents subject to change without notice Salter Brecknell Weighing Products 1000 Armstrong Drive Fairmont MN 56031 Tel 800 637 0529 Tel 507 238 8702 Fax 507 238 8271 E mail s...

Page 2: ...l communi cation port The scale can be configured to display its divisions in two separate modes non NTEP mode and NTEP mode The scale is shipped from the factory configured in non NTEP mode The NTEP...

Page 3: ...The scale is now ready for use ON OFF SWITCH DC POWER JACK RS 232 PORT ID TAG CALIBRATION SWITCH COVER PC 150 Back Panel The female 9 Pin RS 232 connector is used to interface to a PC Shown at right a...

Page 4: ...the displayed weight is within 0 25 divisions of true zero NET Indicates that the scale is displaying net weight GROSS Indicates that the scale is displaying gross weight TARE Indicates that the tare...

Page 5: ...S Toggles between Gross and Net weight display if a tare has been established TARE Used to establish the weight on the platform as the tare weight You cannot establish zero or negative gross weight as...

Page 6: ...off to the unit 2 While holding down the lb kg key turn the power back on 3 When the display shows A1 the unit is in User Menu mode and you can release the lb kg key Shown at right are the direc tion...

Page 7: ...Serial Data Format Selects the data format to be transmitted via the serial port to a printer or computer 0 Consolidated Controls Format 1 Toledo 8213 Format 2 NCI 3825Format 3 Transcell Technology F...

Page 8: ...Shown at right are the di rectional and SET key assign ments SET SETUP MENU CHART F1 Display Mode F6 Zero Range F9 Calib Unit 0 1 F11 Display Check F12 Calibration Press ZERO key to begin Press ZERO...

Page 9: ...one level begins the test sequence Press ZERO key to begin sequence F12 Calibration Places scale into the calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZER...

Page 10: ...y pressing the PRINT key or the TARE key 5 After setting the exact value press the NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then freez...

Page 11: ...0 6 7 Segment Green LED OVER CAPACITY ANNUNCIATION 102 of Full Scale Capacity OPERATING TEMPERATURE RANGE 32 F to 104 F 0 C to 40 C PLATTER SIZE 13 x 13 330 mm x 330 mm POWER SOURCE AC Adapter 12VDC 5...

Page 12: ...al Select the proper settings for the following using the scale s settings Baud Rate Data Bits Parity Stop Bits 1 4 Enable the CAPS LOCK key on your PC 5 This test verifies that the scale is transmitt...

Page 13: ...necessary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to overloading LO Warmup Mode The scale has been turned on without the platter install...

Reviews: