Sage 7937 Operating Instructions Manual Download Page 7

E230169

Medical equipment with respect to electric shock, fire and mechanical  
hazards only, in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.1. 

Consult Instructions for Use

Protective Earth (Ground)

TECHNICAL SPECIFICATIONS

Product Code

7937 (US Configuration)

7937-X (EU configuration)

Input Rating

120 V AC / 60 Hz / 7 AMP

240 V AC / 50 Hz / 3.5 AMP

Target Temperature (B Configuration)

Adjustable between 100°F and 125°F (38°C - 52°C)

Adjustable between 100°F and 125°F (38°C - 52°C)

Target Temperature (C Configuration)

Adjustable between 100°F and 125°F (38°C - 52°C)

Adjustable between 100°F and 125°F (38°C - 52°C)

Default Temperature Setting
(B Configuration)

125°F (52°C)

125°F (52°C)

Default Temperature Setting
(C Configuration)

125°F (52°C)

125°F (52°C)

Ambient Operating Range

60°F - 80°F (16°C - 27°C)

60°F - 80°F (16°C - 27°C)

Relative Humidity Operating Range

50 - 70%

50 - 70%

Operating Pressure

70 kPa - 106 kPa

70 kPa - 106 kPa

Outer Dimensions (H x W x D)

16.88” x 14.69” x 15.25” (429mm x 373mm x 387mm)

16.88” x 14.69” x 15.25” (429mm x 373mm x 387mm)

Net Weight (N.W.) 
Gross Weight (G.W.)

N.W.: 29.5 lb. (13.4 kg)
G.W.: 34 lb.  (15.5 kg)

N.W.: 29.5 lb. (13.4 kg)
G.W.: 34 lb.  (15.5 kg)

Mass with Max Working Load  
(fully loaded unit)

36.8 lb. (16.7 kg)

36.8 lb. (16.7 kg)

Mass with Max Working Load  
(fully loaded cart and unit)

117.3 lb. (53.3 kg)

117.3 lb. (53.3 kg)

Classification

Class I

Class I

Degree of Protection Against Ingress  
of Water

IPXO

IPXO

Mode of Operation

Suitable for Continuous Operation

Suitable for Continuous Operation

Storage / Transport Temperature

-22 - 149°F (-30 - 65°C)

-22 - 149°F (-30 - 65°C)

Storage / Transport Relative Humidity

10% to 90%

10% to 90%

Storage / Transport Atmospheric  
Pressure Range

500 to 1060 hPa

500 to 1060 hPa

Standards - UL / EN / CSA / IEC 60601-1

UL 60601-1 Medical Electrical Equipment

UL 60601-1 Medical Electrical Equipment

UL Classification

IEC 60601-1, 2nd Edition

IEC 60601-1, 2nd Edition

Third Party Evaluation and Testing

IEC 60601-1, 3rd Edition

IEC 60601-1, 3rd Edition

Characteristics

Powder Coated Steel Exterior / Plastic Door and Window / 
Steel Front Flange / Aluminum Interior and Shelves

Powder Coated Steel Exterior / Plastic Door and Window / 
Steel Front Flange / Aluminum Interior and Shelves

Fuse Type

T10AH, 250V

T10AH, 250V

Disposal

Follow national requirements

Follow national requirements

Refer to accompanying documents

Refer to accompanying documents

Protective earth ground terminal

Protective earth ground terminal

Applied Parts

No applied parts

No applied parts

Caution

Equipment not suitable for use in the presence of  
flammable anesthetics

Equipment not suitable for use in the presence of  
flammable anesthetics

- 7 -

Summary of Contents for 7937

Page 1: ...Instructions Model Number 7937 REF 7937 X For warming premoistened cloth and hair care products manufactured by Sage Products LLC PLEASE READ AND UNDERSTAND THESE INSTRUCTIONS COMPLETELY PRIOR TO OPER...

Page 2: ...from heated surfaces 15 The warmer is to be serviced only by qualified service personnel If there are no external or user serviceable fuses contact Sage Products for repair or adjustment For external...

Page 3: ...may be appropriate to cut down on the number of slots filled with that product code and restock them with one of your higher usage products If patient census is low and more than a few slots frequent...

Page 4: ...shock by providing an escape wire for the electric current This appliance is equipped with a cord having a grounded wire with a grounded plug The plug must be plugged into an outlet that is properly...

Page 5: ...indicator counts down If no faults are found LED indicators will turn off if no packages are present in warmer slots If packages are present the LED indicators for the filled slots will indicate NOT R...

Page 6: ...TURE and press the SELECT key below the display 3 When prompted enter the following key sequence UP ARROW DOWN ARROW UP ARROW 4 Using the UP DOWN arrow keys select the desired temperature for the cont...

Page 7: ...6 7 kg Mass with Max Working Load fully loaded cart and unit 117 3 lb 53 3 kg 117 3 lb 53 3 kg Classification Class I Class I Degree of Protection Against Ingress of Water IPXO IPXO Mode of Operation...

Page 8: ...Products LLC 2014 Rev 000 SAVE THESE INSTRUCTIONS FOR PROPER USE OF YOUR WARMER READ ALL SAFETY CAUTIONS AND OPERATING INSTRUCTIONS http www sageproducts com products warmers microwave warmers cfm Man...

Reviews: