2. The warmer must be on a SEPARATE CIRCUIT. No other appliance should share the circuit with the
warmer. If it does, the branch circuit fuse may blow, or the circuit breaker may trip, or Comfort Personal
Cleansing
®
products may warm slower than the times recommended.
3. The VOLTAGE used must be the same as specified on the warmer (115V). Using a higher voltage may result in a
fire or other accident causing warmer damage. Using a lower voltage will cause slow heating. Sage Products is
NOT
responsible for damage resulting from the use of the warmer with other than specified voltage.
General Use
1.
DO NOT
tamper with or make any adjustments to the control panel, housings, safety switches, or any other part
of the Comfort Personal Cleansing Warmer. Repairs should only be done by a qualified service person. Contact
Sage Products for authorized service.
2. If a fire occurs in the warmer,
LEAVE DOOR(S) CLOSED
. Disconnect the power cord, or shut off power at the
fuse or circuit breaker panel.
3.
DO NOT
store flammable materials next to, on top of, in or near the warmer. They could be a fire hazard.
4. Your warmer is equipped with a protective device designed to turn the unit off in the event of overheating. The
warmer will remain inoperable until it has cooled.
CAUTION:
If there is any indication cap or washcloths are excessively hot, do not use. Remember, gloves diminish
your sensitivity to heat. Closely monitor heat level with infants, unresponsive or skin sensitive individuals. Caps/Baths
are for individual use only
.
This warmer is to be used SOLELY in connection with the use of Comfort Personal Cleansing products.
The warmer has been designed and safety-tested to be used exclusively with Comfort Personal Cleansing products.
ANY other use of the warmer by the facility, including but not limited to use with other bathing products, can result
in overheating, fire or other hazardous conditions and is expressly forbidden. Refer to individual packages of Comfort
Personal Cleansing products for recommended heating times.
SAFETY PRECAUTIONS
INSTRUCTIONS FOR USE
Examine
the Comfort Personal Cleansing
®
Warmer before plugging in power cord. If any part appears damaged, do not
use the warmer. Contact your Sage Products Representative.
1.
To install warmer,
position unit on a sturdy, flat surface, such as a counter top or Comfort Personal Cleansing
®
cart. For proper grounding, plug cord into AC receptacle marked “hospital only” or “hospital grade.” Turn power
switch on front panel to “on” position. The red light in the power switch should illuminate when switched to the
“on” position.
When power is turned on, the warmer goes through the following self-diagnostic tests:
•
All red “NOT READY” indicators flash three times.
•
All green “READY” indicators flash three times.
•
All green “TAKE FIRST” indicators flash three times.
•
Temperature indicator counts down.