background image

2.  The warmer must be on a SEPARATE CIRCUIT. No other appliance should share the circuit with the 

warmer. If it does, the branch circuit fuse may blow, or the circuit breaker may trip, or Comfort Personal 
Cleansing

®

products may warm slower than the times recommended.

3.  The VOLTAGE used must be the same as specified on the warmer (115V). Using a higher voltage may result in a 

fire or other accident causing warmer damage. Using a lower voltage will cause slow heating. Sage Products is 

NOT

responsible for damage resulting from the use of the warmer with other than specified voltage. 

General Use

1. 

DO NOT 

tamper with or make any adjustments to the control panel, housings, safety switches, or any other part 

of the Comfort Personal Cleansing Warmer. Repairs should only be done by a qualified service person. Contact 
Sage Products for authorized service.

2.  If a fire occurs in the warmer, 

LEAVE DOOR(S) CLOSED

. Disconnect the power cord, or shut off power at the 

fuse or circuit breaker panel.

3. 

DO NOT 

store flammable materials next to, on top of, in or near the warmer. They could be a fire hazard.

4.  Your warmer is equipped with a protective device designed to turn the unit off in the event of overheating. The 

warmer will remain inoperable until it has cooled.

CAUTION:

If there is any indication cap or washcloths are excessively hot, do not use. Remember, gloves diminish

your sensitivity to heat. Closely monitor heat level with infants, unresponsive or skin sensitive individuals. Caps/Baths
are for individual use only

.

This warmer is to be used SOLELY in connection with the use of Comfort Personal Cleansing products. 
The warmer has been designed and safety-tested to be used exclusively with Comfort Personal Cleansing products.
ANY other use of the warmer by the facility, including but not limited to use with other bathing products, can result
in overheating, fire or other hazardous conditions and is expressly forbidden. Refer to individual packages of Comfort
Personal Cleansing products for recommended heating times.

SAFETY PRECAUTIONS

INSTRUCTIONS FOR USE

Examine 

the Comfort Personal Cleansing

®

Warmer before plugging in power cord. If any part appears damaged, do not

use the warmer. Contact your Sage Products Representative.

1. 

To install warmer,

position unit on a sturdy, flat surface, such as a counter top or Comfort Personal Cleansing

®

cart. For proper grounding, plug cord into AC receptacle marked “hospital only” or “hospital grade.” Turn power 
switch on front panel to “on” position. The red light in the power switch should illuminate when switched to the 
“on” position.

When power is turned on, the warmer goes through the following self-diagnostic tests:

All red “NOT READY” indicators flash three times.

All green “READY” indicators flash three times.

All green “TAKE FIRST” indicators flash three times.

Temperature indicator counts down.

Summary of Contents for Comfort 7944

Page 1: ...Shock Type B Applied Parts Classification Class I Y9891B Sage Products Inc 2003 Degree of Protection Against Ingress of Water Ordinary Mode of Operation Suitable for Continuous Operation Storage Tempe...

Page 2: ...4 14 count Model 7945 28 count Please read and understand these instructions completely prior to operating the Comfort Personal Cleansing Warmer For use with Comfort Personal Cleansing brand bathing i...

Page 3: ...near a swimming pool 8 DO NOT use this warmer for any material other than Comfort Personal Cleansing products in accordance with instructions on the label 9 DO NOT let the cord hang over edge of tabl...

Page 4: ...the warmer long enough to reach the target temperature and is ready for use 3 TAKE FIRST The blinking green arrow and message that reads TAKE FIRST indicates the package in this slot has been in the w...

Page 5: ...e electrical outlet d Be certain electrical outlet is not blocked by the warmer Power cord must be able to be unplugged without moving the warmer Grounding Instructions This appliance must be grounded...

Page 6: ...sively hot do not use Remember gloves diminish your sensitivity to heat Closely monitor heat level with infants unresponsive or skin sensitive individuals Caps Baths are for individual use only This w...

Page 7: ...emove fuses If the warmer stops running check the fuses on the back side of the cabinet s exterior If a fuse is blown replace it with another fuse of identical rating If the new replacement fuse blows...

Page 8: ...umns and center column must be removed to allow cleaning of the shelves and cabinet interior Remove the Phillips head screw located at the base of each column Gently lower the column to disengage from...