background image

 

DOCUMENT NO:  M27   ISSUE:  NO: 07   ISSUE DATE: 27/06/2018 

 

 

 

 

INSTRUCTION MANUAL

 

AIRONE DS Range of 

Semi-Mobile Ducted Fume Cupboards

 

 

 
 

 
 
 

SAFELAB SYSTEMS 

Airone Building ● 8 Beaufighter Road ● Weston-Super-Mare ● BS24 8EE 

Telephone: 01934 421340 

 Fax: 01934 641569 

 E-mail: [email protected] 

WWW.SAFELAB.CO.UK 

 

Summary of Contents for AIRONE DS

Page 1: ...6 2018 INSTRUCTION MANUAL AIRONE DS Range of Semi Mobile Ducted Fume Cupboards SAFELAB SYSTEMS Airone Building 8 Beaufighter Road Weston Super Mare BS24 8EE Telephone 01934 421340 Fax 01934 641569 E m...

Page 2: ...e 4 Principles of Operation Page 5 Installation and Assembly Instructions Page 6 7 Specification Page 8 9 Operating Instructions Page 10 Wiring Diagram Page 11 Calibration Page 12 13 14 Maintenance an...

Page 3: ...ice Engineers or their authorised agents Details of Service Contracts Programmes along with information on these and other products are available on request from SAFELAB SYSTEMS LTD Airone Building 8...

Page 4: ...an air contaminants are effectively scavenged from the work surface by the low level slot created by the back baffle Lighter than air contaminants are drawn upwards to the extract plenum created by th...

Page 5: ...ter on the 1200DS 250mm diameter on the 1500DS 3 Connect to the power supply ensure the extract duct system fan is running and the cabinet is ready for commissioning See Calibration instructions on pa...

Page 6: ...handle The work tray is white polypropylene as standard Sash Operation The vertical sliding laminated glass sash of the AIRONE DS fume cupboard is suspended and counterweighted providing added advanta...

Page 7: ...ng 840 590 1040 590 1340 590 Weight Packed 120kgs TBC TBC Access clearance height 1 915mm SPECIFICATION DIAGRAM 1 Lower storage area 2 Vertical laminated glass sash 3 Laminated glass side rear windows...

Page 8: ...ely 5 minutes prior to use 5 Always use appropriate PPE personal protection equipment when using the Fume Cupboard 6 When working in the cupboard try and ensure that the sliding sash is in its lowest...

Page 9: ...DOCUMENT NO M27 ISSUE NO 07 ISSUE DATE 27 06 18 9 OPERATING INSTRUCTIONS CONTINUED GUIDE TO POSITIONING LOCATIONS FOR THE AIRONE RS MOBILE FUME CUPBOARD...

Page 10: ...DOCUMENT NO M27 ISSUE NO 07 ISSUE DATE 27 06 2018 10 WIRING DIAGRAM...

Page 11: ...rated 100mm diameter vane anemometer 6 Press the Mute Button once and release 7 The LEDs will stop alternating off and on The alarm point is now stored 8 Restore the airflow velocity to what it was pr...

Page 12: ...hazard to the operator We recommend that a service programme be arranged with Safelab Systems ltd Regular maintenance by our qualified personnel will ensure safe running of your equipment and also en...

Page 13: ...orrections from the air flow meter calibration chart Look at the table and repeat any reading which seems to be very different from the general pattern Record the average of this and the previous read...

Page 14: ...above Average m s Drop 10 from Do filters PASS Is Fume flow Any deterioration or damage observed Fail if face velocity Initials of Tester a b c d e f g h i Before Check Inward 0 3 m s If a drop of 10...

Page 15: ...hat the cabinet is switched on during any cleaning procedure and that suitable protective clothing face mask gloves and safety glasses is worn All surfaces should be cleaned with a mild detergent solu...

Page 16: ...DOCUMENT NO M27 ISSUE NO 07 ISSUE DATE 27 06 2018 16 Airone DS Range...