13 Probes and Biopsy
118
Basic User Manual
Table 13-2
Recommended Method for High-level Disinfection
Disinfectant
Manufacturer
Active
Ingredient
Level for Active
Ingredient
Contact
Type
Contact
Period
Cidex™ Activated
Dialdehyde Solution
J&J
Glutaraldehyde
2.4%
Soak
45-50 minutes
Resert XL HLD
STERIS
H
2
O
2
2.0%
Soak
8 minutes
Tristel Duo for
Ultrasound
Tristel
Chlorine
dioxide
0.02%
Wipe
0.5 minutes
Tristel Sporicidal
Wipe* (* part of
the Tristel Trio
Wipes System)
Tristel
Chlorine
dioxide
0.02%
Wipe
0.5 minutes
13�1�5
Sterilizing the Probe
You should sterilize the intraoperative, biopsy use probe or any probe that comes into contact with blood after each
use.
Perform the steps as follows:
1. Clean the probe.
For details, refer to Section 13.1.3 Cleaning the Probe.
2. Sterilize the intraoperative probe as shown in Table 13-3.
Or select the recommended disinfectant (see Appendix D.3 Disinfectant) to sterilize the probe by following the
instruction provided by disinfectant’s manufacturer.
3. Rinse the probe with running sterile water to completely remove the residual sterilant.
4. Dry the probe with a lint-free soft dry cloth.
Table 13-3
Recommended Method for Sterilization
Disinfectant
Manufacturer
Active
Ingredient
Level of Active
Ingredient
Contact
Type
Contact Period
Cidex™ Activated
Dialdehyde Solution
J&J
Glutaraldehyde
2.4%
Soak
10 hours
13�1�6
Disinfecting and Sterilizing the Probe Cable
!
To avoid potential disease transmission, you should wear sterile medical gloves and protective goggles during
disinfection or serialization.
Do not immerse the probe cable in alcohol or isopropyl alcohol at any concentration for a long period of time.
Avoid any contact with iodine or phenols. Otherwise, it will accelerate the aging of the cable and decrease the
service time of the probe, or even result in cable damages.
■
Daily Disinfection
Perform the following steps to clean the probe cable.
1. Use a lint-free soft dry cloth dampened with mild soapy water to wipe the probe cable.
Summary of Contents for EVUS 8
Page 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Page 10: ...This page is intentionally left blank...
Page 18: ...This page is intentionally left blank...
Page 62: ...This page is intentionally left blank...
Page 88: ...This page is intentionally left blank...
Page 92: ...This page is intentionally left blank...
Page 112: ...This page is intentionally left blank...
Page 122: ...This page is intentionally left blank...
Page 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...