21
S
ae
bo
Stim Micro
Treatment Intensity
Two buttons on the device control intensity: Increase (
▲
) and
Decrease (
▼
). See Figure 16. The intensity defaults to 5 when the
device is powered on or when the batteries have been removed and
replaced.
Press the Increase button (
▲
) to set the protocol and intensity
level. The “P” will stop flashing and as you continue to press the
button, the intensity will increase and the LCD will display the
numeric value. The intensity will increase in increments of 5, up to
the maximum intensity of 100. To decrease the intensity, press the Decrease (▼) button.
How much intensity do I need?
Press the Increase button (
▲
) until you feel a light tingling sensation. Next, reduce the
sensation by pressing the decrease button (
▼
) until you no longer feel the tingling.
According to research, the optimum therapeutic level of stimulation should be just below
“sensory” level. In other words, try to reduce the stimulation to the point where you no
longer feel the electricity or tingling sensation.
Note: The Increase and Decrease buttons lock out after 20 seconds of non-use so that
stimulation cannot be inadvertently changed during treatment. To reactivate the intensity
controls, press and release the Power button. Intensity controls will remain active for 20
seconds after the last button was pressed.
Treatment Intensity