5-10
Graphic Settings
This page is currently inactive.
5-11
Clinical Setup
The following items could be set in this menu:
Smart nebulizer
Humidification type
Compliance compensation
O2 Sensor
EXH Flow Sensor
Figure 5-15 Clinical Setup
Smart nebulizer
Activation, deactivation and duration of nebulizer operation can be specified in this section.
By activating the nebulizer from this menu, the nebulizer activates for the time specified by
the user and the related message is displayed on the screen. After this period, it will be
deactivated automatically.
Available options for the duration of nebulizer activity are: 1 min to 480 min. Default setting
is 20 min. Any change in the nebulization time setting will be saved until the next change.
Nebulizer is not activated in inspiratory flow (air and oxygen) less than 11L/min and “Neb.
Not Usable” alarm message will be displayed.