EMC
Guidance and manufacturer's declaration – electromagnetic emissions
The Dena Electrocardiograph is intended for use in the electromagnetic environment specified
below. The customer or the user of the Dena should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Dena uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Dena is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Use only the recommended manufacturer accessory .Using the
accessory other than in relevant chapter may cause to increase the
EMISSION or decrease the IMMUNITY of system.
Measurements can be affected by mobile and RF communications
equipment. It should be assured that the Electrocardiograph is used
in the electromagnetic environment specified
To prevent EMC effect on the Electrocardiograph, the system
should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, the equipment
should be observed to verify normal operation in the
configuration in which it will be used.
Do not use cellular phone in the vicinity of this equipment. High
result in strong level of electromagnetic radiation emitted from such