VW2106 Vibrating Wire Readout Manual
RST Instruments Ltd.
Page 12
1.
Scroll to the
Monitor Settings
screen using the up/down arrows and press
Enter
.
2.
Select the
Import from Loc.
option and scroll to the desired location using the
up/down arrows and press
Enter
when complete.
The location label will appear at the top of the screen to indicate the current location
in use.
NOTE:
T
HE DEVICE WILL DISPLAY THE MESSAGE
“N
O
L
OCATIONS
”
IF NO LOCATIONS
EXIST IN THE READOUT
’
S MEMORY
.
3.7
C
REATING
C
USTOM
P
ARAMETERS
The following instructions outline the steps to set up custom parameters.
1.
Scroll to the
Monitor Settings
screen using the up/down arrows and press
Enter
.
2.
Select the
Custom Setup
option and press
Enter.
3.
Select the number of sensors connected and press
Enter
. Most devices will be
set to connect to one sensor per terminal strip (Figure 3-11). The expansion
connector is automatically enabled when more than one sensor is selected.
F
IGURE
3-11
O
NE SENSOR CONNECTED
4.
Select the appropriate sweep frequency and press
Enter
. Refer to Table 3-3 for
sweep frequencies corresponding to sensor type.
T
ABLE
3-3
S
WEEP
F
REQUENCIES
Sweep
Frequency
Sensor Type
A
450-6000Hz
Wide Sweep
B (Default)
1200-3550Hz
Piezometer, Strain Gauge, Borehole Stressmeter, Jointmeter,
Crackmeter, Displacement, Settlement, Temperature, Load
Cells