background image

RS Stock No.712-6033

Instruction Manual

RS-1316

Dual Data Logger Thermometer

FR

EN

IT

DE

ES

TYPE

MX

AVG

- MN

INTERVAL

min:sec

hour:min

T2

T1

T1

T2

C
F
K

KJTERSN

Summary of Contents for RS-1316

Page 1: ...RS Stock No 712 6033 Instruction Manual RS 1316 Dual Data Logger Thermometer FR EN IT DE ES TYPE MX AVG MN INTERVAL min sec hour min T2 T1 T1 T2 C F K KJTERSN...

Page 2: ...OF CONTENTS EN 10 15 10 Version No 00 TABLE OF CONTENTS TITLE PAGE 1 INTRODUCTION 1 2 SPECIFICATIONS 2 3 FRONT PANEL DESCRIPTION 4 4 OPERATION INSTRUCTIONS 7 5 MAINTENANCE 13 6 RECALIBRATION PROCEDUR...

Page 3: ...ollows the international temperature scale of 1990 ITS 90 Read the following information carefully before attempting to operate or service the meter When servicing use only specified replacement parts...

Page 4: ...to 1767 0 C 35 to 1999 9 F S type 2 0 to 1767 0 C 35 to 1999 9 F Display effective resolution J K T E and N type 0 1 C F K R and S type 1 0 C F K 0 1 C F K only for reference Measurement accuracy J K...

Page 5: ...ry capacity 98 sets Continuity data logging capacity 4100 sets 2 2 General specifications Power supply Qty 6 AAA batteries Battery life approx 80 hours Auto power off 30 minutes If no key is pressed L...

Page 6: ...15 10 Version No 00 3 FRONT PANEL DESCRIPTION T1 T2 14 10 12 13 1 5 4 2 3 7 9 8 15 6 11 C D A B 16 TYPE MX AVG MN INTERVAL min sec hour min T2 T1 T1 T2 C F K KJTERSN T1 KJTERSN TYPE hour min min sec...

Page 7: ...to turn the meter on or off 4 MX MN key Press MX MN key to step through the maximum minimum and average readings Press MX MN key for 2 seconds to exit MX MN mode 5 T1 T2 T1 T2 key Press T1 T2 T1 T2 ke...

Page 8: ...d time in the MX MN mode 10 TYPE key Press TYPE key to select the thermocouple type K J E T R S or N 11 READ key Press READ key to show manual memory logged readings Press again to exit this mode 12 C...

Page 9: ...ess key to turn on the thermometer Plug the thermocouple s into the thermocouple input s as required If no thermocouple is plugged into the selected input or the thermocouple is open circuit the appro...

Page 10: ...isplay between hour min and min sec on the LCD display The elapsed time since entering reading mode or the time at which the MAX MIN or AVG value occurred will appear on the time display Press T1 T2 T...

Page 11: ...imum reading plus elapsed time T2 Averaging reading plus averaging time TYPE min sec K MX MN AVG C T1 T2 present reading T1 T2 Maximum reading plus elapsed time T1 T2 Minimum reading plus elapsed time...

Page 12: ...and all data stored in memory will be cleared 4 4 To Trigger One by One Datalogging Press the MEM key once and one set of readings will be stored in memory The LCD display shows MEM and a memory loca...

Page 13: ...release it The LCD display will show INTERVAL MEM and interval time Press or key until the display shows the required logging interval 3 to 255 seconds and then press key to select Press the MEM key...

Page 14: ...wnloaded to a PC It cannot be read using the READ function to show the data the on display 4 6 How to disable auto power off function The meter will automatically turn off if no key press occurs for 3...

Page 15: ...cally cleans the case with a damp cloth and mild detergent Do not use abrasives or solvents Clean and dry as required 5 2 Battery Replacement When the LCD display shows BT the battery has insufficient...

Page 16: ...Version No 00 6 RECALIBRATION PROCEDURE The thermometer should be calibrated once a year to ensure its continued accuracy Contact RS Components for further details of calibration service The address...